Filling procedure with
funnel filler
w WARNING
Step 1
Remove the filler cap by turning it
counter-clockwise. Ensure that the
drain plug is closed by tightening it
with the tool end of the filler cap.
Step 2
Verify that the anesthetic agent is the
same as that specified on the
vaporizer front label. Observe the
agent level through the sight glass
indicator on the side of the filler body.
Pour the agent slowly into the filling
port, as illustrated on Fig. 4-3, until
¥
the level reaches the
mark. The
level may decrease slightly as the
wicks absorb the agent. To help
prevent overfill, ensure the agent level
is at or below the fill line.
Step 3
Replace and tighten the filler cap to
minimize leaks.
1175-0013-000
Before filling a vaporizer equipped with a funnel filler, turn the cap slowly
to allow any pressure to gradually vent.
Ensure that the drain plug screw, located on the lower front of the
vaporizer, is correctly tightened to help prevent loss of liquid agent.
Figure 4-3 • Filling a vaporizer that incorporates a funnel filler
Figure 4-4 • Replacing the filler cap
4 Operating Instructions
4-5