Filling procedure with
Quik-Fil
w w w w WARNING
Step 1
Remove the protective cap from the
anesthetic agent bottle filler, checking
that the bottle and filler mechanism
are not damaged.
Step 2
Remove the filler cap. Insert the bottle
nozzle into the filler block. Rotate the
bottle to align the bottle filler nozzle
keys with the index slots in the filler
block as illustrated on Fig. 4-11.
Step 3
Press the agent bottle fully into the
vaporizer filler block. Allow the liquid
to flow into the vaporizer until the
¥
maximum level mark
is reached.
Pay particular attention to the level in
the sight glass and the air return
bubbles flowing into the bottle.
Step 4
Release the bottle when the vaporizer
is full and the continuous stream of
bubbles ceases. To help prevent
overfill, ensure the agent level is at or
below the fill line.
Step 5
Remove the bottle from the vaporizer
filler. Replace the filler cap and the
cap on the agent bottle. Ensure that
the filler cap is tightened to help
minimize leaks.
1175-0013-000
Ensure that the drain plug screw, located on the lower front of the
vaporizer, is correctly tightened to help prevent loss of liquid agent.
3
2
1
1. Filler cap
2. Index slot
3. Nozzle key
Figure 4-11 • Filling a vaporizer with a Quik-Fil
Figure 4-12 • Replacing the filler cap
4 Operating Instructions
4-9