Draining Procedure With Funnel Filler - Datex-Ohmeda Tec 7 User's Reference Manual

Table of Contents

Advertisement

Tec 7 Vaporizer
Draining procedure
with funnel filler
w w w w CAUTION
Step 1
Remove the filler cap. Insert the tool
end of the cap into the drain plug
below the filling port on the filler body
as illustrated in Fig. 4-5.
Step 2
Position a properly marked container
under the drain spout.
Step 3
Unscrew, but do not remove, the drain
plug to allow the vaporizer contents to
pour from the drain spout into the
container.
Step 4
After draining is complete, tighten the
drain plug to help minimize leaks.
Step 5
Replace and tighten the filler cap to
help minimize leaks.
4-6
The vaporizer must only be drained into a properly marked container.
Do not allow the container to become completely full during draining
procedures.
Figure 4-5 • Draining a vaporizer with a funnel filler
Figure 4-6 • Replacing the filler cap