Filling Procedure With Easy-Fil - Datex-Ohmeda Tec 7 User's Reference Manual

Table of Contents

Advertisement

Filling procedure
with Easy-Fil
w w w w WARNING
Step 1
Align the notches on the bottle
adapter to the bottle collar and tighten
the adapter onto the agent bottle.
Step 2
Remove the filler cap. Align the bottle
adapter keys with the index slots in the
filler block as illustrated in Fig. 4-7.
Insert the bottle nozzle into the filler
block.
Step 3
Press the agent bottle fully into the
vaporizer filler block. Allow the liquid
to flow into the vaporizer until the
maximum level mark ¥ is reached. To
help prevent overfill, ensure the agent
level is at or below the fill line. Pay
particular attention to the level in the
sight glass and the air return bubbles
flowing into the bottle.
Step 4
Release the bottle when the vaporizer
is full and the continuous stream of
bubbles ceases.
Step 5
Remove the bottle from the vaporizer
filler. Replace the filler cap and the
cap on the agent bottle. Ensure that
the filler cap is tightened to help
minimize leaks.
1175-0013-000
Ensure that the drain plug screw, located on the lower front of the
vaporizer, is correctly tightened to help prevent loss of liquid agent.
The filling system consists of three elements:
the bottle collar
the bottle adapter
the filler block
The vaporizer must only be filled using the correct agent specific filling system
3
2
1
1. Filler cap
2. Index slot
3. Bottle adapter
Figure 4-7 • Filling a vaporizer with Easy-Fil
Figure 4-8 • Replacing the filler cap
4 Operating Instructions
.
4-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents