Checking the vaporizer for correct mounting

w w w w WARNING

Removing the vaporizer from a manifold

Step 1
Set the dial to
.
OFF
If the dial is not completely turned to
the
position the vaporizer cannot
OFF
be released from the manifold.
3-4
To help ensure correct operation, do not use a vaporizer that is either
visibly out of line on the manifold or that can be lifted off the manifold
when the locking lever is in the locked position.
If more than one vaporizer is fitted, visually check to make sure that the tops of the
vaporizers are level. If the vaporizer is visibly out of line, perform steps 2 and 3 as
described in Removing the vaporizer from a manifold and remount it correctly.
When the vaporizer appears to be level and the locking lever is in the locked position,
attempt to lift the vaporizer straight up from the manifold. If the vaporizer can be lifted
off the manifold, it is not correctly mounted. Remount the vaporizer (see Vaporizer
mounting procedure ).
Verify that the interlock rods are in alignment by making sure that only one vaporizer at
a time can be turned ON.
Check the anesthesia system for leaks in accordance with the relevant User's
Reference Manual with the vaporizer dial turned to 0% and then repeat the check with
the vaporizer dial turned to
Figure 3-6 • Setting the dial
.