Mounting the vaporizer

Step 1
Set the dial to
.
Step 2
Unlock the locking lever.
Turn the lever counter-clockwise.
Make sure the lever releases.
Step 3
Prepare the manifold.
Remove any plugs fitted to the
vaporizer interlock block ports.
Verify that each manifold port
valve O-ring is intact. If
necessary, remove the existing
O-rings and fit one new O-ring to
each port valve, as described in
the relevant anesthesia system
User's Reference Manual.
Replacement O-rings are
supplied with each vaporizer.
3-2
Figure 3-1 • Setting the concentration dial
Figure 3-2 • Unlocking the locking lever
1. Vaporizer Interlock Block Port - ensure plugs removed
2. Replace Manifold Port Valve O-ring, if necessary
Figure 3-3 • Readying the manifold
1
2
1175-0013-000