Vaporizer mounting procedure

w w w w WARNING
1175-0013-000
3 Setup and Mounting Procedure
The vaporizer is designed to be used on Selectatec Series Mounted Manifolds. The
vaporizer can be installed on other Selectatec Manifolds but the interlock system is
designed to function on Selectatec Series Mounted Manifolds only.
Mounting a Tec 7 Vaporizer on a Selectatec 7 Compatibility Block is not
recommended.
Do not lift or support the vaporizer by holding the control dial. Handle the
vaporizer with care at all times.
Before mounting a vaporizer onto the Selectatec Series manifold,
ensure that each manifold port valve O-ring is intact and that there is no
foreign matter around the mating surfaces. A damaged O-ring and/or
foreign matter around the mating surfaces can cause leaks.
Earlier versions of the Selectatec Series Mounted Manifold that provide
mounting positions for three vaporizers require that if only two vaporizers
are fitted, then the center position must be occupied. If the center
position is not occupied, the interlock that helps ensure that only one
vaporizer at a time can be turned ON is ineffective.
Do not use a vaporizer if the liquid level decreases below the minimum
level.
Before using a vaporizer allow it to attain the ambient temperature of the
location in which it has to be used.