Performance; Weight And Dimensions - Datex-Ohmeda Tec 7 User's Reference Manual

Table of Contents

Advertisement

Tec 7 Vaporizer

Performance

Weight and dimensions

7-2
Accuracy at 5 liters/
Vaporizer 5%: ± 0.25% of delivered agent or ± 15% dial setting
min O
21 ± 2 °C
(whichever is greater)
2
Vaporizer 8%: ± 0.4% of delivered agent or ± 15% dial setting
(whichever is greater)
Liquid capacity
To fully charge a vaporizer with dry wicks: 300 ml (nominal)
Retained by wick system: 75 ml (nominal)
To fill from minimum to maximum mark:
5% Vaporizer: 170 ml (nominal)
8% Vaporizer (Sevotec):
137 ml (nominal)
Flow resistance at
10 - 15 cm H
5 liters/minute of
O
at 21 ± 2 °C
2
Operating
18 °C to 35 °C (64 °F to 95°F)
temperature range
Storage temperature
–40 °C to 65 °C (–40 °F to 149°F)
range
Note: Protect the vaporizer packaging from condensation.
Weight
7 kg
Depth
210 mm
Width
110 mm
Height
250 mm
O with Vaporizer setting ON +0%
2
15 lb 6 oz (empty)
8.25 inches
4.375 inches
9.875 inches (in the unlocked position)
1175-0013-000