Datex-Ohmeda Field Service Support Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel.
Field Operations Unit or to an Datex-Ohmeda Authorised Distributor. Additional copies of this manual, quoting Tec 5 Continuous Flow Vaporizer O & M Manual Part No. 1105-0100-000, can be requested from a local Datex-Ohmeda Field Operations Unit or from an Datex- Ohmeda Authorised Distributor.
Servicing Policy During initial installation of a Tec 5 vaporizer, the Next Service Date label located on the vaporizer dial must be completed by the Technician/Engineer/Customer responsible for the installation. Datex-Ohmeda recommends that all Tec 5 Vaporizers are serviced every three years, irrespective of conditions of use.
The vaporizer must be locked on to the manifold before it can be turned ON. 2. Only one vaporizer at a time can be turned ON when two or more Tec 4, Tec 5 or Tec 6 vaporizers are fitted on a Selectatec Series Mounted Manifold.
Selectatec Manifolds but the interlock system is designed to function on Selectatec Series Mounted Manifolds only. Mounting a Tec 5 vaporizer on a Selectatec 7 Compatibility Block is not recommended. Warning: It is a requirement of European Standard prEN 740 - Anaesthetic Workstations And Their Modules, that the gas monitoring device referred to at 3.1 of this manual is in operating condition,...
Fig. 3. PUSH DOWN TO FULL EXTENT LOCKED TURN LOCKING LOCKING LEVER LEVER DIAL DIAL RELEASE RELEASE Fig. 3 Locking The Vaporizer On To A Manifold Tec 5 Vaporizer Page 5 O & M Manual Part No. 1105-0100-000 August 1999...
Note that it is not possible to turn ON the vaporizer if an adjacent Tec 4, Tec 5 or Tec 6 vaporizer is turned The vaporizer should not be used between OFF and the first graduation mark. To avoid inadvertent...
Turn the vaporizer dial clockwise to the OFF position. FILLER CAP HEXAGON SIGHT GLASS DRAIN PLUG Fig. 6 Filling A Vaporizer Which Incorporates A Screw Cap Filler Tec 5 Vaporizer Page 7 O & M Manual Part No. 1105-0100-000 August 1999...
5. Replace the filler cap and tighten it to minimise any possibility of leaks. FILLING PORT DRAIN PLUG HEXAGONAL FILLER CAP DRAIN SPOUT CONTAINER Fig. 7 Draining A Vaporizer Which Incorporates A Screw Cap Filler Page 8 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
PORT CLAMP VALVE LEVER LEVER 021A T5 OM AGENT SPECIFIC BOTTLE COLLAR AGENT SPECIFIC BOTTLE COLLAR Fig. 8 Filling A Vaporizer Which Incorporates A Keyed Filler Tec 5 Vaporizer Page 9 O & M Manual Part No. 1105-0100-000 August 1999...
BOTTLE ADAPTOR LEVER LOCKING CLAMP LEVER LOCKING CLAMP AGENT SPECIFIC LEVER BOTTLE COLLAR 022A T5 OM Fig. 9 Draining A Vaporizer Which Incorporates A Keyed Filler Page 10 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
5. Withdraw the bottle from the vaporizer filler and replace the filler cap (2) and the cap on the agent bottle. Ensure that the filler cap is tightened to minimise any possibility of leaks. Fig. 10 Filling A Vaporizer Which Incorporates A Quik-Fil™ Filler Tec 5 Vaporizer Page 11 O & M Manual Part No. 1105-0100-000...
Note: If further drain funnels are required they should be obtained from the local Abbott Laboratories representative. Fig. 11 Draining A Vaporizer Which Incorporates A Quik-Fil™ Filler Page 12 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
5.1.2 Every Three Years Datex-Ohmeda recommends that all Tec 5 Vaporizers are serviced every three years, irrespective of conditions of use. The vaporizer should be serviced at an Datex-Ohmeda authorised Service Centre.
Return the vaporizer to a Service Centre stating, if possible, the type of contaminant used. Repairs Repairs must only be performed by the Manufacturer's Service Representatives. Page 14 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
CONTROL DIAL CONTROL DIAL RELEASE VAPORIZER LOCKING LEVER PORT VALVE ACTUATING SPINDLE EXTENSION PORT PORT VALVES VALVES SELECTATEC SERIES MOUNTED MANIFOLD Fig. 12 Vaporizer Interlock Mechanism Tec 5 Vaporizer Page 15 O & M Manual Part No. 1105-0100-000 August 1999...
EXTENDED TO PREVENT THE POSSIBILITY OF TURNING ACTUATING ON VAPORIZER B SPINDLE Fig. 14 Two Vaporizers Locked On To The Manifold - Vaporizer A Turned ON Page 16 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
Tec 4, Tec 5 and Tec 6 vaporizers. Warning: Connecting any vaporizer other than a Datex-Ohmeda Tec 4, Tec 5 or Tec 6 vaporizer to the Selectatec Series Mounted Manifold defeats the interlock system. This action results in the ability to administer anaesthetic agent mixtures, possibly resulting in patient injury.
Conversely, if the temperature of the vaporizer increases, the thermostat opens, the by-pass flow through the thermostat increases and therefore the vaporizing chamber flow decreases. Thus the output of the vaporizer remains relatively constant under conditions of changing temperature. Page 18 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
6.4.1 General The output concentration of the Tec 5 vaporizer is regulated by the 'variable flow-split' method described in the following text and illustrated on Fig. 16 and 17: A total flow of fresh gas from upstream flowmeters enters the vaporizer at where it is immediately split into two streams.
GAS OUT FRESH GAS BY-PASS FRESH GAS OUT IPPV COMPENSATING ASSEMBLY WICK ASSEMBLY VAPORIZING CHAMBER THERMOSTAT 015A T5 OM Fig. 17 Tec 5 Vaporizer Gas Flow Page 20 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
Effect of Flowrate at 21 ± 2° C with oxygen flowing 002A T5 OM Flowrate (litres/minute oxygen) Effect of Temperature at 5 litres/minute oxygen flow 003A T5 OM Temperature ° C Tec 5 Vaporizer Page 21 O & M Manual Part No. 1105-0100-000 August 1999...
Effect of Flowrate at 21 ± 2° C with oxygen flowing 004A T5 OM Flowrate (litres/minute oxygen) Effect of Temperature at 5 litres/minute oxygen flow 005A T5 OM Temperature ° C Page 22 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
Effect of Flowrate at 21 ± 2° C with oxygen flowing 006A T5 OM Flowrate (litres/minute oxygen) Effect of Temperature at 5 litres/minute oxygen flow 007A T5 OM Temperature ° C Tec 5 Vaporizer Page 23 O & M Manual Part No. 1105-0100-000 August 1999...
Effect of Flowrate at 21 ± 2° C with oxygen flowing 008A T5 OM Flowrate (litres/minute oxygen) Effect of Temperature at 5 litres/minute oxygen flow 009A T5 OM Temperature ° C Page 24 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
Effect of Flowrate at 21 ± 2° C with oxygen flowing 024A T5 OM Flowrate (litres/minute oxygen) Effect of Temperature at 5 litres/minute oxygen flow 023A T5 OM Temperature ° C Tec 5 Vaporizer Page 25 O & M Manual Part No. 1105-0100-000 August 1999...
% v/v must be changed in inverse proportion to the barometric pressure. The vaporizer automatically makes this % v/v change and for practical clinical purposes the effects of barometric pressure can be ignored. Page 26 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
ON. This is a normal machine characteristic and is caused by residual vapour left in the machine from previous use. Tec 5 Vaporizer Page 27 O & M Manual Part No. 1105-0100-000...
In order to help to achieve the reliability and consistency of performance of Tec 5 Vaporizers, Datex- Ohmeda use closely specified test conditions, test methods and detailed protocol in conjunction with training, experience and quality auditing systems.
Nevertheless, it would be unusual for the measured liquid consumption to vary by more than 25% of the preceding values. Tec 5 Vaporizer Page 29 O & M Manual Part No. 1105-0100-000...
±0.25% of delivered agent or Vaporizer 5% ±15% of dial setting (whichever is greater). ±0.4% of delivered agent or Vaporizer 8% ±15% of dial setting (whichever is greater). Page 30 Tec 5 Vaporizer August 1999 O & M Manual Part No. 1105-0100-000...
The foregoing warranties shall not apply if the Product has been repaired or serviced other than by Datex-Ohmeda or other than in accordance with written instructions provided by Datex-Ohmeda, or altered by anyone other than Datex-Ohmeda, or if the Product has been subject to abuse, misuse, negligence, or accident.