8.

Typical Errors (Error Codes)

Display
Error code
Error 1
EC,E11
Error 2
Error 3
Error 6
EC,E16
Error 7
EC,E17
Error 8
Error 9
Con Err
door Err
S/N Error
CAL E
EC,E20
-CAL E
EC,E21
Weighing
display
E
Description and possible countermeasure
Stability error
Weighing value is unstable and "zero display", "sensitivity
adjustment", "calibration test", etc. cannot be executed. Check
around the pan. Refer to the separate instruction manual listed in
"1.1. Detailed Manuals". Improve the environment of the installation
location (vibration, drafts, static electricity, etc.).
Out of the setting range
The value to be set exceeds the setting range. Set again within the
setting range.
Malfunction of the internal memory element of the balance
If this error continues to be displayed, repair is necessary. Please
contact your local dealer for repair.
Internal mass error
Applying the internal mass does not yield a change in the mass value
as specified. Confirm that there is nothing on the pan and perform the
operation from the beginning.
Internal mass error
The internal mass application mechanism does not function properly.
Perform the operation from the beginning.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary.
Please contact your local dealer for repair.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary.
Please contact your local dealer for repair.
Weighing unit error
Data from the weighing unit has not been received. Check the
connection between the weighing unit and the display unit, and turn
the power off and then on again.
Breeze break unit error
Communication with the breeze break is not possible. Reconnect the
cable between the breeze break unit and the weighing unit.
Serial number error
The serial numbers of the weighing unit and the display unit do not
match. Reconnect with the correct combination.
Calibration weight error (Positive value)
The calibration weight is too heavy. Check around the pan. Check the
calibration mass value.
Calibration weight error (Negative value)
The calibration weight is too light. Check around the pan. Check the
calibration mass value
Overload error
A sample beyond the balance weighing capacity has been placed on
the pan. Remove the object from the pan.