DMTA-20006-01EN, Rev. B, February 2014
3.
If you need to replace the probe:
a) Using the T6 Torx screw driver, remove the sleeve clamp screw, and then slide
the protective sleeve and sleeve clamp away from the tail (see Figure 2-1 on
page 22).
Sleeve clamp screw
b) Remove the sleeve clamp completely.
c)
Extract the probe cable, couplant tube, and encoder cable (encoder side tail
only) from the protective sleeve.
2.2

Installing the Probe

To install the probe
1.
Insert the probe cable, couplant tube, and encoder cable (encoder side tail only) in
the cable-management portion of the tail.
2.
Using the yellow loom installation tool, insert the protective sleeve over the cables
and the tube as follows (see Figure 2-2 on page 23):
a) Starting at the scanner end, close the tool over the bundle with the pointed
shoe portion pointing away from the scanner (see illustration A).
b) Slide the protective sleeve over the pointed shoe portion of the tool so that it
spreads open, follows the tool, and starts to wrap itself around the bundle
(see illustrations C and D).
c)
While holding the portion of the protective sleeve that is already around the
bundle, slide the tool away from the scanner so that the remainder of the
protective sleeve wraps around the bundle (see illustration D).
22
Chapter 2

Figure 2-1 Removing the sleeve clamp