to use the electrodes on clean and dry skin. When using the electrodes, follow
the instructions given in the manual and on the package of the electrodes.
Use only single-patient electrodes, supplied exclusively by the manufacturer,
and take care to avoid the exchange of electrodes between different users.
MIO-CARE has been tested and guaranteed for the use only with the
electrodes supplied by the manufacturer;
to use ONLY accessories supplied by device manufacturer. Only use battery
chargers supplied by the manufacturer; the use of battery chargers not
supplied by the manufacturer will free the same from any responsibility
related to damage to the equipment or user and will expose the user to risks
such as short circuits and fire.
It is forbidden:
to use the device in the presence of patient monitoring equipment, of
electrosurgical (possible bruises and burns) or shortwave or microwave
therapy equipment or other equipment that sends electrical impulses into
the body and in general in combination to other medical devices, since it
could cause problems to the stimulator;
to use the device close to flammable substances/gas/explosives, in
environments with high concentrations of oxygen, with aerosol-therapy
devices or in wet environments (use of the device is prohibited in bathroom
or shower areas or while showering/bathing);
to use the device by persons known to be unsound-minded, or suffering from
sensibility disorders, permanently or temporarily disabled unless assisted by
qualified personnel (e.g. a doctor or therapist); by persons younger than 15
years old or not adequately educated about the device use by an adult
person;
to use the device in presence of signs of deterioration of the device itself,
cables and/or electrodes: please contact the dealer or the manufacturer
following the instructions given in the paragraph Support. Control carefully
the integrity of the device before each use;
to position the electrodes in such a way that the current crosses the heart
area (e.g. a black electrode on the chest and a red electrode on the shoulder
blade); however, electrodes can be positioned along the muscular fascia of
the heart area, as used for pectoral strengthening. Danger of heart arrythmia;
to position the electrodes close to the eyes; make sure that the current
delivered does not cross the eyeball (one electrode diametrically opposite to
the other in relation to the eye); keep a distance of at least 3 cm from the
eyeball;