I-Tech MAG1000 User Manual

I-Tech MAG1000 User Manual

Magnetotherapy model
Hide thumbs Also See for MAG1000:

Advertisement

Quick Links

USER MANUAL
Magnetotherapy model
MAG1000

Advertisement

Table of Contents
loading

Summary of Contents for I-Tech MAG1000

  • Page 1 USER MANUAL Magnetotherapy model MAG1000...
  • Page 3: Table Of Contents

    INDEX INDEX TECHNICAL SPECIFICATIONS ANUFACTURER ECLARATION OF CONFORMITY LASSIFICATIONS URPOSE AND SCOPE ECHINICAL FEATURES EVICE AND COMMANDS DESCRIPTION ABELLING Package content HOW TO USE NTRODUCTION TO TECHNOLOGY ONTROINDICATIONS Side effects ARNINGS ATIENT PREPARATION POSITIONING OF THE THERAPEUTIC BELT NSTRUCTIONS FOR USE DEVICE CARE AINTENANCE ROUBLESHOOTING...
  • Page 4: Technical Specifications

    I.A.C.E.R. S.r.l Via S.Pertini 24/A – 30030 Martellago (Ve), Italy herewith declares under its own responsibility, that the product MAG1000 UMDNS Code: 12415 has been designed and manufactured according to the European Medical Device Directive 93/4/EEC (transposed in Italy by the D.Lgs. 46/97), as modified by the Directive 2007/47/EC (D.Lgs.37/2010) and further...
  • Page 5: Classifications

    • degeneration of locomotor apparatus • sprains • periarthritis • benign lesions and muscular tears • tendonitis MAG1000 is particularly suitable for the treatment and the care of the osteoporosis and all the pathologies on bony tissue. IACER Srl MNPG249-03...
  • Page 6: Techinical Features

    The patient population intended for magnetotherapy treatment using the MAG1000 device includes patients of both sexes, men and women, of age (unless otherwise indicated by medical doctors). For further details, please refer to the Contraindications section. Techinical features...
  • Page 7: Device And Commands Description

    Device and commands description Label (retro) ON/OFF Key Start/pause key Time set key Programs selection key Field intensity key IACER Srl MNPG249-03...
  • Page 8: Labelling

    Battery status indication PROGRAM P-01 Selected program for therapy PULSE RATE 8Hz Selected pulse rate Magnetic field intensity level selected: INTENSITY 1 - low, 2 - medium, 3 - high TIMER 45’00’’ Therapy time (minutes and seconds) Working channel: A - channel1, B- channel2 Labelling The label on the side is placed on the back of...
  • Page 9: Package Content

    15° from its normal functioning position. Power supply (DC 6.8V / 0.3A). A / B (CH1 / CH2) channel 1/2. Non ionizing Package content The Mag1000 package contains: n°1 MAG1000 device; • • n°1 elastic therapeutic belt; n°1 battery pack (inside the device);...
  • Page 10: How To Use

    (PEMF, carrying frequency from 20 to 30 MHz, with frequencies up to 5.000 Hz). The MAG1000 has been designed to exploit these fields, in fact it generates an electric signal with a carrier frequency of 27mhz, which is modulated with specific frequencies and therefore is able to soothe the anatomical site affected by pain.
  • Page 11: Controindications

    Controindications Patient in pregnancy, tuberculosis, juvenile diabetes, viral (in acute phase) illnesses, mycosis, cardiopathic subjects, serious arrhythmias or pacemaker carriers, children, metallic prosthesis carriers, acute infections, epileptics (different medical prescriptions excepted). The functioning of some electric implantable devices, such as pacemakers, could be compromised during a treatment with a short-wave therapy device.
  • Page 12 WARNING! During therapy it is possible to hear a light hiss coming from MAG1000: such operation is normal and must not arouse worry. The manufacturer shall be deemed to be responsible for the performance,...
  • Page 13: Patient Preparation: Positioning Of The Therapeutic Belt

    Wrist Elbow Cervical Instructions for use Follow the instructions for the correct operation of the MAG1000: 1) Switch on MAG1000 pressing ON/OFF key 2) Check battery status: if the icon , on the upper side of the display is flashing, proceed to recharge the unit following the instructions (see chapter Battery charging.
  • Page 14 6) MAG1000 is now ready for use: the display shows information about the program in use, the working frequency (not adjustable), the adjusted therapy time and field intensity (adjustable during the treatment). 7) Press key: the display will show the remaining therapy time, while it will be possible to modify field intensity during the whole treatment, pressing more times IN.
  • Page 15 IACER S.r.l. suggest the therapy time as shown on the table and pre-setted on MAG1000. However, the user can adjust the time as he prefers. MAG1000 uses therapy time values, working frequency values and field intensity values coming from scientific and medical literature, as result of well-known sperimentations and clinical evaluations (Barker - Lunt 1983, Bassett –...
  • Page 16: Device Care

    Store the device in a cool, well-ventilated place. Do not store heavy objects on the device. It is recommended to switch off MAG1000 at the end of each treatment and to remove the cables from the connectors. MAG1000 should be kept in the...
  • Page 17: Troubleshooting

    PROGRAM. To proceed with the charging, switch off MAG1000 pressing connect the battery charger to the plug on the upper side of MAG1000 MAG 1000 is equipped with a software protection that switch off the equipment when the battery charger is connected. Carry out two 5-hour charging cycles (maximum time set for charging) to obtain a full charge of the batteries, as it takes at least 8/10 hours.
  • Page 18: Replacing The Battery

    • Contact the manufacture or authorized dealer for original spare part; • Open the battery compartment on the back side of MAG1000, disconnect the red/black wire from connector and remove the exhausted battery; • Insert new battery and connect the red/black wire to the plug;...
  • Page 19: Warranty

    Warranty IACER Srl guarantees a warranty period from the purchasing date for MAG1000 device, unless information contained in this manual regarding installation, use and maintenance is strictly adhered. The wearing parts (batteries) are not included in the warranty, unless of visible manufacturing defects.
  • Page 20: Assistance

    Warranty conditions 1) Should assistance be needed, enclose the purchasing receipt when sending the device to the manufacturer. 2) The warranty period is valid only on the electronic parts. The warranty will be granted by the shop or directly by the manufacturer.
  • Page 21: Spare Parts

    Spare parts The manufacturer shall make available the original spare parts for the equipment at any time. To request them: I.A.C.E.R. S.r.l. Via S. Pertini, 24/a • 30030 Martellago (VE) Tel. 041.5401356 • Fax 041.5402684 For the purpose of maintaining the warranty, the functionality and safety of the product it is recommended to use only original spare parts supplied by the manufacturer (also consult the Warnings paragraph).
  • Page 22 Guidance and manufacturer's declaration – ELECTROMAGNETIC EMISSIONS – FOR ALL DEVICES AND SYSTEMS The MAG1000 device is expected to operate in the electromagnetic environment below specified. The customer or user of the MAG1000 device must ensure that it is used in such environment. Electromagnetic environment –...
  • Page 23 Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR ALL DEVICES AND SYSTEMS MAG1000 is intended for use in the electromagnetic environment specified below. The user or operator of MAG1000 should assure that is used in such environment. Electromagnetic Test level...
  • Page 24 Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR ALL DEVICES AND SYSTEMS MAG1000 is intended for use in the electromagnetic environment specified below. The user or operator of MAG1000 should assure that is used in such environment. Electromagnetic Test level...
  • Page 25 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MAG1000 is used exceeds the applicable RF compliance level above, the MAG1000 should be observed to verify normal operation.
  • Page 26 MAG1000 that are not life-supporting MAG1000 is intended for the use in an electromagnetic environment in which radiated RF disturbances are controlled. The user or the operator of MAG1000 can help prevent electromagnetic interferences by maintaining a minimum distance...

This manual is also suitable for:

12415

Table of Contents