USER MANUAL
Magnetotherapy model
MAG700

Advertisement

Table of Contents
loading

Summary of Contents for I-Tech MAG700

  • Page 1 USER MANUAL Magnetotherapy model MAG700...
  • Page 3 INDEX INDEX TECHNICAL SPECIFICATIONS ANUFACTURER ECLARATION OF CONFORMITY LASSIFICATIONS URPOSE AND SCOPE ECHNICAL FEATURES EVICE AND COMMANDS DESCRIPTION ABELLING Package content HOW TO USE NTRODUCTION TO TECHNOLOGY ONTRAINDICATIONS AND SIDE EFFECTS ARNINGS ATIENT PREPARATION POSITIONING OF THE THERAPEUTIC BELT MAIN APPLICATIONS AND SUGGESTIONS ETUP AND PROGRAMS INSTRUCTIONS Connections and power on...
  • Page 4: Declaration Of Conformity

    MAG2000 which includes the following models MAG700, MAG2000, MAG2000 Premium and MAG2000 PLUS UMDNS Code: 12415 has been designed and manufactured according to the European Medical Device Directive 93/4/EEC (transposed in Italy by the D.Lgs. 46/97), as modified by the Directive 2007/47/EC (D.Lgs.37/2010) and further...
  • Page 5: Purpose And Scope

    MAG700 is particularly recommended for the treatment and care of osteoporosis and all pathologies borne by bone tissues. Thanks to the high intensity of the magnetic field that can generate, MAG700 is particularly indicated in the treatment of bone fractures even in the presence of rigid bandages or plaster.
  • Page 6: Technical Features

    The patient population intended for magnetotherapy treatment using the MAG700 device includes patients of both sexes, men and women, of age (unless otherwise indicated by medical doctors). For further details, please refer to the Contraindications section. Technical features...
  • Page 7 Device and commands description Label (retro) Power button and return to the program choice menu. OK key, confirmation button. Selection/increment key. Selection/decrement key. During the therapy a green light will turn on. IACER Srl MNPG69-09...
  • Page 8 1. ON/OFF button 2. Power supply entrance 15V/1.2A 3. CH1 connector 1 Labelling The label on the side is placed on the back of the device IACER Srl MNPG69-09...
  • Page 9 The label is placed on protective envelope. Symbol Description Manufacturer logo Certification of the product issued by the notified body N° 0068. Manufacturer Manufacturing date (YYYY-MM) Attention, consult operating instructions. Product subject to WEEE regulations concerning separate waste collection of electronic equipment. Class II equipment.
  • Page 10 Only domestic usage. Not protected against liquid entrance, keep dry. Power supply symbol. Generic warning symbol: pay attention to the danger indicated Package content The MAG700 package contains: • N°1 MAG700 device; • N°1 power supply (cable approx. 1.5m); •...
  • Page 11: Contraindications And Side Effects

    BMD (Bone Mass Density). MAG700 is a single channel device, it’s very easy and intuitive to use with a high prices/performance ratio. Contraindications and side effects...
  • Page 12 • do not damage the applicator by acting on the connecting wire, and do not wrap the wire around the applicator or the appliance; • avoid the use of the system to persons not adequately educated by reading the manual. Keep out of the reach of children, animals and pests;...
  • Page 13 • the electrical system of the environment in which MAG700 is inserted complies with national laws; • the appliance is used in strict accordance with the operating instructions contained in this manual.
  • Page 14 Patient preparation: positioning of the therapeutic belt, main applications and suggestions Here below a list of main positions for the therapeutic belt and for the solenoids couple. Wrap the belt around the area to be treated (or position the belt on the area, for example in vertebral column treatment).
  • Page 15: List Of Programs

    INSERTED inside the envelope of protection! Main menu The MAG700 device is equipped with 11 programs with preset and user- modifiable values according to your needs. The first 8 programs are associated with the treatment of specific pathologies with preset...
  • Page 16 MAG700 uses therapy time values, working frequency values and field intensity values coming from scientific and medical literature, as result of well-known experimentations and clinical evaluations (Barker - Lunt 1983, Bassett –...
  • Page 17 3. Press the button by highlighting the magnet icon in the lower left then press OK. 4. The device will begin the treatment, displaying the magnet icon with the magnetic field flow on the display. The green light below the display alerts you to the current therapy. 5.
  • Page 18 continue the treatment. During the pause time green led turns off until the treatment restarts. ATTENTION: it’s possible to get out from the treatment at any time by pressing once key, the screen will display the basic settings (step 2). By pressing again key the screen will display programs menu (step 1).
  • Page 19 4. To start the therapy and continue as in the previous paragraph from step 2, press the button and then OK to select the magnet icon and confirm the start of the treatment. Green light indicates that therapy is running. At the end of therapy, the screen will display automatically the menu programs.
  • Page 20: Functioning Control

    If you are experiencing any problems or if you need any further information, please contact the manufacturer immediately. FUNCTIONING CONTROL MAG700 is equipped with a magnet (small ring or metal or metal/plastic disc) in order to control the device functioning. Control procedures: 1.
  • Page 21: Carriage And Storage

    CARRIAGE AND STORAGE Carriage precautions MAG700 is a portable device, so it does not need any particular carriage precautions. However, we recommend putting away MAG700 and its accessories in their own bag after every treatment, and store everything inside the packaging box.
  • Page 22: Warranty

    Warranty IACER Srl guarantees a warranty period from the purchasing date for MAG700 device, unless information contained in this manual regarding installation, use and maintenance is strictly adhered. The wearing parts (applicators’ fabric as well as elastic velcro closure of the same) are not included in the warranty, unless of visible manufacturing defects.
  • Page 23 Warranty conditions 1) Should assistance be needed, enclose the purchasing receipt when sending the device to the manufacturer. 2) The warranty period is valid only on the electronic parts. The warranty will be granted by the shop or directly by the manufacturer.
  • Page 24 (also consult the Warnings paragraph). Interference and elctromagnetic compatibility tables The MAG700 equipment has been designed and manufactured according to the TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY legislation EN 60601-1-2:2015 with the aim of providing adequate protection from harmful interference when installed in homes and health establishments.
  • Page 25 In particular, in order to prevent any interference problems, it is advisable to operate any therapy device sufficiently distant from critical equipment to monitor patient's vital functions and to use caution in therapeutic applications on patients carrying cardiac stimulators. For more details consult the compatibility tables at the end of the manual. IACER Srl MNPG69-09...
  • Page 26: Electromagnetic Compatibility Tables

    ELECTROMAGNETIC COMPATIBILITY TABLES Guidance and manufacturer's declaration – ELECTROMAGNETIC EMISSIONS – FOR ALL DEVICES AND SYSTEMS MAG2000 family is expected to operate in the electromagnetic environment below specified. The customer or user of the MAG2000 family must ensure that it is used in such environment. Electromagnetic environment –...
  • Page 27 Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR ALL DEVICES AND SYSTEMS MAG2000 is intended for use in the electromagnetic environment specified below. The user or operator of MAG2000 should assure that is used in such environment. Electromagnetic Test level Compliance Immunity test environment –...
  • Page 28 Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY FOR DEVICES AND SYSTEMS THAT ARE NOT OF VIABLE FUNCTION The MAG2000 Family is expected to operate in the electromagnetic environment below specified. The user or operator of the MAG2000 family must ensure that it is used in this environment.
  • Page 29 Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY FOR DEVICES AND SYSTEMS THAT ARE NOT OF VIABLE FUNCTION At 80 MHz and 800 MHz the separation distance for the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
  • Page 30 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. MAG700. All rights reserved. MAG700 and logos are owned by IACER and are registered. Edition: MNPG69-09 of the January 31...