Fluke 902 Calibration Information Manual page 9

Hvac clamp meter
Hide thumbs Also See for 902:
Table of Contents

Advertisement

Test
(Switch Position)
V
AC Voltage
V
DC Voltage
P
Continuity
P
Resistance
G
DC Current
**F
Temperature
L
Capacitance
K
*
AC Current
*Test using the 5500A Coil
**Use K-Type TC Extension Wire with 80CK, at both ends, for connection between 5520A and 80AK/Meter
Table 2. Performance Tests
5520A Output
600 V/60 Hz
10 V/60 Hz
600 V dc
0 V dc
-600 V dc
25 Ω
9500 Ω
1500 Ω
950 Ω
0 Ω
200 µA dc
10 µA dc
0 µA dc
-200 µA dc
-40 °C
0 °C
400 °C
1 µF
90 µF
200 µF
1 mF
0.4 A, 60 Hz
12 A, 60 Hz
12 A, 50Hz
HVAC Clamp Meter
Performance Tests
Meter Response
Lower Limit
593.5
606.5
9.4
10.6
593.5
606.5
-0.5
0.5
-606.5
-593.5
Beeper on
9353
9647
1473
1527
935.3
964.7
-0.5
0.5
197.5
202.5
9.4
10.6
-0.5
0.5
-202.5
-197.5
-43.5
-36.5
-0.8
0.8
395.2
404.8
0.8
1.2
88.1
91.9
195
205
979
1021
19.1
20.9
587.5
612.5
587.5
612.5
Upper Limit
9

Advertisement

Table of Contents
loading

Table of Contents