Error Messages - HP ESG Series User Manual

Signal generators
Hide thumbs Also See for ESG Series:
Table of Contents

Advertisement

Troubleshooting

Error Messages

Error Messages
If an error condition occurs in the signal generator, it is reported to both the front panel
display error queue and the SCPI (remote interface) error queue. These two queues are
viewed and managed separately.
NOTE
When there is an unviewed message in the front panel error queue, the
annunciator appears on the signal generator's display.
Table 3-1 Characteristics of the Error Queues
Characteristic
Capacity (#errors)
Overflow Handling
Viewing Entries
Clearing the Queue
Permanent Errors
(errors that must be
resolved, for example:
unlock, ovencold, and
hi/lo)
3-6
Front Panel Display
Error Queue
30
Circular (rotating).
Drops oldest error as new
error comes in.
Press:
,
,
Utility
Error Info
(or
)
Next
Previous
Error
Message
Press:
,
,
Utility
Error Info
Error Queue(s)
Re-reported after queue is
cleared.
HP ESG and HP ESG-D Series Signal Generators
SCPI Remote Interface
Error Queue
30
Linear, first-in/first-out.
Replaces newest error with:
-350,Queue overflow
Use SCPI query
View
SYSTem:ERRor? or
STATus:QUEue?
Power up
Clear
Send a *CLS command
Read last item in the queue
Re-reported after queue is
cleared.