TROUBLESHOOTING

The Compact Turbimeter has internal diagnostics to
advise on some issues as follows :-
Error 100 – Measurement Off-Scale
The reading is above 1050 NTU. Dilute using the
provided PT 512 Dilution Tube and clean water and
re-measure.
Error 101 – Incorrect Standard
The standard inserted is either the incorrect value or is
no
longer
within
an
acceptable
tolerance
for
calibration. Use the correct standard or replace the
expired calibration set.
Error 107 – Blanking Error
The sample cuvette cap is not attached allowing
ambient light into the QuadoptiX™ chamber or the
sample cuvette is not fully inserted. Replace the cap
and read again or fully insert the sample cuvette.
Error 110 – Battery Low
The available power is insufficient to support reliable
analysis. Replace the batteries.
26