Total Suspended Solids - palintest CT12 Manual

Compact turbimeter
Table of Contents

Advertisement

To cease measurement, press 'Back' at any time.
To index a cuvette, insert the sample cuvette with the
orientation mark facing forward.
Systematically rotate the sample cuvette 45° at a time
and record the position of the lowest value. This is the
optimal cuvette alignment for low turbidity measurement.

Total Suspended Solids

This mode will only be active if a site calibration has
been stored within the Turbimeter. Once a calibration
has been created (via the 'Calibration' menu) the site
ID can be used to recall the site specific relationship
between turbidity and suspended solids.
Choose 'Reading' mode, select 'Total Suspended
Solids' and press 'OK'.
Highlight the Site ID for the Site/Sample under
investigation using the Up/Down arrows. The calibration
data will be recalled automatically.
Insert the sample with the orientation mark facing
forward and press 'Read'.
A beep will occur at the start and end of each
measurement cycle.
Result is displayed on the screen in mg/l and stored in
the log with time, date, Operator ID and Sample ID.
16