Total Suspended Solids - Factor - palintest CT12 Manual

Compact turbimeter
Table of Contents

Advertisement

Press 'Select' to enter the concentration of suspended
solids determined externally using the Up/Down buttons.
When data entry is complete press 'Done' and the
slope/intercept will be calculated and shown auto-
matically.
Data points can be edited or deleted as required.
Highlight the data point of interest and select 'Edit'
Choose either 'Edit' or 'Delete' as required. When
complete select 'Done'
Press 'Back' to return to the 'Mode' menu.
Total Suspended Solids – Factor
Select 'Calibration' and press 'OK'
Select 'Total Suspended Solids' and press 'OK'.
If the Sample ID has already been created, use the
Up/Down arrows to highlight and press 'OK' to select.
If a new Sample ID is required, move the highlight to a
blank area and press 'New'.
Create the new Sample ID using the Up/Down buttons
to select characters as describe in the System section.
Select 'Factor' to enter the data entry screen.
Highlight 'Slope' and press 'OK'.
Use the Up/Down buttons to enter the slope calculated
externally. The value is limited between 0.1 and 50.0.
Press 'Done' to return to the 'Factor' screen.