Total Suspended Solids - Correlation - palintest CT12 Manual

Compact turbimeter
Table of Contents

Advertisement

Press 'Back' to leave the calibration process and return
to the Calibration menu.
To cease calibration during the data capture phase
press 'Cancel'.
To cease the calibration process at any step, press
'Back' (calibration data will be discarded).
The time, date and Operator ID are stored in the
calibration log for GLP compliance (see Page 9,
System menu).
Total Suspended Solids – Correlation
Select 'Calibration' and press 'OK'.
Select 'Total Suspended Solids' and press 'OK'.
If the Sample ID has already been created use the
Up/Down arrows to highlight and press 'OK' to select.
If a new Sample ID is required, move the highlight to a
blank area and press 'New'.
Create the new Sample ID using the Up/Down buttons
to select characters as describe in the 'System' section.
Select 'Correlation' to enter the data entry screen.
The default (0, 0) point is always included. This can be
deleted or edited if not required.
Select 'Add' to recall data points from memory for
which a correlated suspended solids value has been
measured.