Average Mode
Choose 'Reading' mode, select 'Average' and press
'OK'.
Select the number of readings required for the
averaging function :-
Short Average – 3 consecutive readings
Medium Average – 6 consecutive readings
Long Average – 12 consecutive readings
Insert the sample with the orientation mark facing
forward and press 'Read'.
The instrument will beep at the start and end of
measurement cycle.
Result is displayed on the screen and stored in the log
with time, date, Operator ID and Sample ID.
Continuous-Capture Mode
Choose 'Reading' mode, select 'Continuous-Capture'
and press 'OK'
Insert the sample with the orientation mark facing
forward and press 'Read'.
A beep will occur at the start and end of each
measurement cycle.
To store data in the log, press 'Capture' (repeat as
required).
15