palintest CT12 Manual page 19

Compact turbimeter
Table of Contents

Advertisement

Calibrating the Compact Turbimeter
SDVB
Select Calibration and press 'OK'.
Choose 'SDVB' and press 'OK'.
Select the lowest (0.05 – 0.15) NTU standard and insert
into the sample chamber with the orientation mark
facing forward.
Use the Up/Down buttons to adjust the value on screen
to that on the standard and press 'Read'.
Repeat the process for the 20 NTU, 100 NTU and 800
NTU standards.
If standards have expired, are inserted in incorrect
sequence or the 100 and 800 NTU standards are not
inverted prior to use Error 101: Incorrect Standard
may be shown.
After the final 800 NTU standard, the screen will show
'Calculating...Please wait' and a countdown will
appear to show time remaining.
On completion of the process the Compact Turbimeter
screen will show 'Successful'.
Press 'Back' to leave the calibration process and return
to the Calibration menu.
To cease calibration during the data capture phase
press 'Cancel'.
To cease the calibration process at any step, press
'Back' (calibration data will be discarded).