C2 Reference Standards
The chosen reference standard is standard A according to DIN 67530/ISO 2813 or the
optional metal standard according to ISO 7668 for high-gloss measurements.
C3 Gloss Measurement
Turn on REFO with key Meas/ON
Use the program key to select program P1
Use the angle selector key to select the viewing angle (the selected angle is shown on the
display)
Place REFO on the sample to measure
Press key Meas/ON
Now the measured value is shown on the display and stored automatically with consecu-
tive numbering.
C4 Difference Gloss Measurement
Difference measurement means measuring against a reference. The measured differ-
ence between reference and sample is the indicated measured value.
Turn on REFO with key Meas/ON
Use the program key to select program P2
Use the angle selector key to select the viewing angle (the selected angle is indicated on
the display)
Place REFO on the reference
Press key Meas/ON to measure the reference. The reference measurement is confirmed
on the display with letter R next to the angle value
Place REFO on the sample to measure
Press key Meas/ON to measure the difference
Now the measured difference value is shown on the display and stored automatically
with consecutive numbering.
C5 Three-Angle Measurement
In three-angle measurements the sample is measured automatically at viewing angles of
20°, 60° and 85°.
Turn on REFO with key Meas/ON
Use the program key to select program P3
Use the angle selector key to select the first viewing angle
Place REFO on the sample to measure
Press key Meas/ON
Wait for the measuring display
Call the other measurings with the angle selector key
7