IKA WiCo T25 Easy Clean Control Operating Instructions Manual page 17

Table of Contents

Advertisement

Safety:
Time Out:
In the "Time Out" menu, the user can set a time limit. This time limit
applies if there is a communication failure between the drive unit and
the WiCo or the communication range is exceeded.
The drive unit will continue to run at the set speed until the set time
has expired. After that the drive unit will continue to run at the set safe
speed (see the "Safe speed" menu).
Note: The initial setting for the time limit is 30 seconds. The user can set
a value of up to 3 minutes for this time limit.
Safe Speed:
In the "Safe Speed" menu, the user can specify a speed that is appropriate
and safe for the dispersing task. The safe speed applies if there is a commu-
nication failure between the drive unit and the WiCo or the communication
range is exceeded.
Note: The default safety speed is 0 rpm and can be set alternatively be-
tween 1000 and 3000 rpm in 100 rpm increments, and is implemented
after expiry of the time limit (see "Time out").
Password:
In the "Password" menu, the user can protect the WiCo settings using a pass-
word (factory setting: 000).
If you forgot your password, please contact our service department.
Settings:
Languages:
The "Languages" option allows the user to select the desired language
by turning and pressing the rotary/push knob. A tick indicates the lan-
guage that is set for the system.
Display:
The "Display" option allows the user to change the background color
and brightness of the working screen.
The user can also select to activate or inactivate the "Firmware Update
Info" screen at the starting of the WiCo. A tick (√) shows that the option
is activated.
Sound:
The "Sound" option allows the user to activate/deactivate the key-press
sound and to set the volume.
Factory Settings:
Select the "Factory Settings" option by turning and pressing the rotary/
push knob. The system requests confirmation to recreate the factory
settings. Pressing the "OK" button resets all the system settings to the
original standard values set at dispatch from the factory (see "Menu
structure").
Communication:
The "Device name" menu option allows the user to edit the device name.
This can relatively identify the paired drive unit.
The "Bluetooth
" option allows the user to activate/deactivate the "Blue-
®
tooth
®
" function. A tick (
) shows that the option is activated.
Information:
The "Information" option offers the user an overview of the most impor-
tant system settings of the drive unit.