7. Keep a record of the results with other hand copies of PM data kept on site.

2-3-3 Chassis Leakage Current Test (Continued)

Date sheet for Chassis Source Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than
the value shown in Table 2-4 below.
MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE
Country
Normal
Condition
USA
N/A
Others
0.1mA
Caution:
When the meter's ground switch is OPEN, do not touch the unit to avoid
possible electric shock.
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Unit
Tester
Power
POLARIT
Y Switch
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
Table 2-3
CURRENT
Open Ground
0.3mA
0.5mA
Table 2-4
Tester
Test 1 Probe
NEUTRA
Connector
L Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
19
LOGIQ A1 BASIC SERVICE MANUAL
Reverse
Polarity
0.3mA
0.5mA
Test 2 Real
Test 3 Mounting screw
Panel
on the handle
5176463-100
Open Neutral
N/A
0.5mA