GE MEDICAL SYSTEMS
REV 2
10. Check if has mechanical problems or keyboard problems.
11. Check the equipment to see if any hardware has missed.
12. Check the work of fan which is used to decrease temperature, and to ensure the air
circulation.

6-3-2 Cleaning

The surface of all parts can use snug humid soft cloth to clean. It is better to use
litmusless soap, but not alcohol generic liquid
System
1. Cleaning CRT filtrate screen.
2. Use vacuum cleaner and scrubbing brush to clean fan and the mainframe of LOGIQ
A1.
3. Before cleaning system, turn off the power supply first, and then start to do the
cleaning.
4. Cleaning the place around mainframe and system shell.
5. Dab of the monitor with the soft humid cloth to dab, if it is too dirty, it is better to
use humid cloth with scour, and then soft clean cloth.
6. Dab keyboard with soft humid cloth.
7. Dab keyboard with dry soft cloth.
8. Dab the Video frequency printer with dry soft cloth.
Note:
Do not spill any liquid to get inside system when cleaning monitor, be carefully
do not abrade monitor.
Probe
Clean probe surface cable and shell examine probe whether damnification, in order
no liquid coming into probe. If find any dilapidation, do not use that probe before
recovered.
1. Wipe off the ultrasound gel with soft cloth or the running water.
2. Wipe off all the eyeable leftover with snug water soft sponge with litmusless soap
cotton yearn or cloth also. Do not use scour or abrasive to clean probe.
3. Wipe off all the soap leftover with distilled water.
4. Dry the probe with soft cloth or put the probe on the air.
LOGIQ A1 BASIC SERVICE MANUAL
dissoluble substance or abrasive.
34
5176463-100