4.7.2
Canadian Registration Number
CRN
The requirements of CRN are met when a Rosemount 2160 CSA-approved
vibrating fork level detector model is configured with 316/316L stainless
steel (1.4401/1.4404) process-wetted parts and either NPT threaded or 2-in.
to 8-in. ASME B16.5 flanged process connections.
4.8
Europe
4.8.1
ATEX Intrinsic Safety (IS)
I1
Summary of product certification
Protection
Certificate
Markings
Safety
instructions
Special Conditions For Safe Use (X)
1. Use with Rosemount SmartPower options P/N 753-9220-0001 only.
2. The surface resistivity of the antenna is greater than 1 GΩ. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned with
solvents or a dry cloth.
3. The Rosemount 2160 enclosure is made of aluminum alloy and given
a protective epoxy coating; however, care should be taken to protect
it from impact or abrasion if located in a Zone 0.
24
0F04227.2C
Intrinsic Safety (IS) for gas atmospheres
Baseefa 09ATEX0253X
II 1 G
Ex ia IIC T5...T2 Ga
IP66
See
Instructions for hazardous area installations (I1 and I7)
Rosemount 2160 Wireless Level Detector