2.2. AC Input/Output Power Wiring

The UFM-ST has two AC connectors; an Input connector (AC Mains) and an Output connector (Load). Each
of these connectors also has a ground terminal.
W A R N I N G : The System Integrator or end user is responsible for determining and meeting all
safety and compliance requirements necessary to integrate the UFM-ST into the final
application.

2.2.1. AC Input Power Wiring

The UFM-ST three-terminal input connector
AC2 terminals are the AC power lines (Internally fused at 10 Amps). The Ground terminal is connected
internally to the external Ground Stud and must be connected to Ground.
Wiring to the UFM-ST filter must use #16 AWG wires or larger and have an insulation rating of at least 300 V.
D A N G E R :  Ground connection must be connected to ground to minimize possibility of
electrical shock hazard.
14
(Figure
2-1) is the AC power (Mains) connector. The AC1 and
AC1 AC2
INPUT
100-240V~
50/60Hz
10A MAX
FIRST CONNECT TO EARTH
HIGH LEAKAGE CURRENT
Figure 2-1:
AC Input Connections