Chapter 5 ________________________________________________________________ Operation
NOTE
VAISALA _______________________________________________________________________ 77
WET
The WET command without a parameter shows the scaling factor that
normalizes the RAINCAP
state to be 1.00. A typical value is close to 0.0015.
DRD WET SCALE 0.00169
The WET ON command is used to set the parameter. The RAINCAP
measuring surfaces must be coated with a wet cloth or immersed in water
for at least 30 seconds, when the WET ON command is given. Make
sure the wet cloth touches only the RAINCAP
The wet scale normalizes the measured signal range (DRY - WET) to
1.00.
HEAT
This test command can be used to test the heater control relays and the
heating elements and power.
The following is an example of the command:
HEAT ON
When you want to stop the heater test and quit the test mode, press ESC.
Do not leave the heater testing on in warm conditions for long periods of
time.
®
signal change from the dry state to the wet
®
sensor.
®