7.6 Linearity calibration
To enter the service mode (verified scale) you have to slide the switch on the PCB in the On position.
No. Action
1
While holding down
display.
2
Release the keys
3
When the scale after a brief wait shifts to weighing mode,
hold down
until the display shows.
4
Hold F until
5
With a empty weighing pan, press first
together and release both
6
on 0 starts blinking
7
Follow the steps at the display
8
After busy appears, the sclae shifts to the weighing mode
9
Turn off the scale
10
Turn on.
Setting the cumulative weight table for linearity adjustment
( ): Cumulative tital of weights on the scale
Capacity
620 g
Display
on 0
0 g
(0 g)
on 1
150 g
(150 g)
on 2
150 g
(300 g)
on 3
150 g
(450 g)
on 4
170 g
(620 g)
Used
100 g x 4
Weights
50 g x 4
20 g x 1
- 19 -
and
, press ON to turn on the
and
3200 g
0 g
0 g
(0 g)
(0 g)
500 g
3 kg
(500 g)
(3 kg)
1000 g
3 kg
(1500 g)
(6 kg)
1000 g
3 kg
(2500 g)
(9 kg)
700 g
3 kg
(3200 g)
(12 kg)
1000 g x 2
2 kg x 4
500 g x 2
1 kg x 4
200 g x 1
Result / display
Calibration mode
FUNC
CAL2
On 0
Zero point calibration
On 1 to on 4
Weighing mode
12000 g