Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
Ensure that there are no objects on the weighing pan.
When the PRINT-key is pressed during the adjusting procedure, [STOP] will be displayed
and adjustment interrupted. The balance returns to weighing mode.
At the models with internal adjustment weight (KERN PNJ) the adjustment with external
weight is not possible.
The following error messages may be displayed during adjustment.
1-Err Wrong adjustment weight (< 50% max)
2-Err Divergence last external adjustment > 1%
3-Err Weighing pan loaded
4-Err Divergence from last internal adjustment > 1%
A-Err Internal adjustment automatics defective
Err 710 Instable environmental conditions