Aqua Stim Water Irrigator Instructions for Use - US
The external water tank should be placed at the same level or up to 24 inches (60 cm) below the irrigator
and water can still be drawn into the irrigator using the internal pump. Do not place the external water tank
higher than the irrigator.
Water is drawn from the external water tank into the irrigator for heating to 30°C or 44°C prior to each
irrigation. Water will remain in the irrigator until it has been emptied by you following the draining
instructions.

2.3 Marking

The following marking can be found on the instrument:
Symbol
or
Do not place the water tank at a position above the irrigator as this may affect
water fill and overflow functions.
Prior to transport, follow the water draining instructions in this manual. Do not
ship the irrigator without draining it as the internal water may damage
components due to freezing and this will void the warranty.
Explanation
Type B applied parts.
Patient
applied parts that are not conductive and can be immediately
released from the patient.
Refer to instruction manual
WEEE (EU-directive)
This symbol indicates that when the end-user wishes to discard this product,
it must be sent to separate collection facilities for recovery and recycling.
Failing to do so may endanger the environment.
Year of manufacture