Environment; External Suction Pump With Power Supply - GE KISS Service Manual

Electrode application system
Table of Contents

Advertisement

Signal connector
Connector to a CAM-14 Acquisition Module
Negative pressure adjustment for pump module
continuously adjustable between 80 and 220 mbar; in position
< 30 mbar
Mechanical data
swivel arm length
vertical swivel range
horizontal swivel range
length of electrode leads
length of negative pressure tube connected to separate pump or ECG Device:
270 cm

Environment

Operation
• temperature between +10°C and +40°C
• relative humidity between 25 and 95%, non-condensing
• atmospheric pressure between 700 and 1060 hPa
Storage and transport
• temperature between -30°C and +70°C
• relative humidity between 10 and 95%, non-condensing
• atmospheric pressure between 500 and 1060 hPa
Weight
electrode distributor with 14 electrodes approx. 950 g

External Suction Pump with Power Supply

Power supply
from the power line; design in compliance with IEC 60601-1/EN 60601, protection
class II; for rated line voltages from 100 to 240 VAC, 50 to 60 Hz (without switchover)
input operating voltage range
input current
output voltage
output current
22745010H
75 to 125 cm (extendible)
45°
360°
limbs 140 cm
chest 120 cm
100 to 240 VAC
700 mA-350 mA
12 V
2.5 A
KISS™ / KISS™ Multilead
Specifications
negative pressure
49

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kiss multileadKiss-10Kiss-12

Table of Contents