GE KISS Service Manual
GE KISS Service Manual

GE KISS Service Manual

Electrode application system
Table of Contents

Advertisement

GE Healthcare
KISS™ / KISS™ Multilead
Electrode Application System
Service Manual
22745010 Revision H
English
© 1996 - 2017 General Electric Company.
All Rights Reserved.

Advertisement

Table of Contents
loading

Summary of Contents for GE KISS

  • Page 1 GE Healthcare KISS™ / KISS™ Multilead Electrode Application System Service Manual 22745010 Revision H English © 1996 - 2017 General Electric Company. All Rights Reserved.
  • Page 2 Publication Information The information in this manual applies only to KISS /KISS Mlutilead . It does not apply to earlier product versions. Due to continuing product innovation, specifications in this manual are subject to change without notice. KISS, CASE and MAC are trademarks owned by GE Medical Systems Information Technologies, Inc., a General Electric Company going to market as GE Healthcare.
  • Page 3 Any unauthorized attempt to repair equipment under warranty voids that warranty. • It is the user’s responsibility to report the need for service to GE Healthcare or to one of their authorized agents. • Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
  • Page 4 Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a pochopili jeho obsah. • V případě nedodržování této varování může dojít k poranění pracovníka prodejního servisu, obslužného personálu nebo pacientů vlivem elektrického proudu, respektive vlivem mechanických či jiných rizik. 22745010H KISS™ / KISS™ Multilead January 2017...
  • Page 5 Ne pas tenter d'intervenir sur les équipements tant que le manuel technique n'a pas été consulté et compris. • Le non-respect de cet avertissement peut entraîner chez le technicien, l'opérateur ou le patient des blessures dues à des dangers électriques, mécaniques ou autres. 22745010H KISS™ / KISS™ Multilead...
  • Page 6 Ez a szerviz kézikönyv kizárólag angol nyelven érhető el. FIGYELMEZTETÉS • (HU) Ha a vevő szerviz ellátója angoltól eltérő nyelvre tart igényt, akkor a vevő felelőssége a fordítás elkészíttetése. • Ne próbálja elkezdeni használni a berendezést, amíg a szerviz kézikönyvben leírtakat nem értelmezték és értették meg.
  • Page 7 • Nemėginkite atlikti įrangos techninės priežiūros, nebent atsižvelgėte į šį eksploatavimo vadovą ir jį supratote. • Jei neatkreipsite dėmesio į šį perspėjimą, galimi sužalojimai dėl elektros šoko, mechaninių ar kitų paslaugų tiekėjui, operatoriui ar pacientui. 22745010H KISS™ / KISS™ Multilead...
  • Page 8 Este manual de assistência técnica só se encontra disponível em inglês. (PT-BR) • Se o serviço de assistência técnica do cliente não for GE, e precisar de outro idioma, será da responsabilidade do cliente fornecer os serviços de tradução. •...
  • Page 9 Eğer müşteri teknisyeni bu kılavuzu İngilizce dışında bir başka lisandan talep ederse, bunu tercüme ettirmek müşteriye düşer. • Servis kılavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz. • Bu uyarıya uyulmaması, elektrik, mekanik veya diğer tehlikelerden dolayı teknisyen, operatör veya hastanın yaralanmasına yol açabilir. 22745010H KISS™ / KISS™ Multilead...
  • Page 10 Không tuân thủ những cảnh báo này có thể dẫn đến các tổn thương cho người thực hiện sửa chữa, người vận hành hay bệnh nhân, do sốc điện, các rủi ro về cơ khí hay các rủi ro khác. 22745010H KISS™ / KISS™ Multilead January 2017...
  • Page 11: Table Of Contents

    Testing Electrical Function (KISS™ Multilead) ........
  • Page 12 1. Test Plate (p/n 30344490) ..............45 2. Y-Adapter (p/n 30344489) ..............45 3. Adapter Extension with Syringe (p/n 2081397-001) ......45 4. Dummy Plug for KISS™ (p/n 30344488) ..........46 Specifications KISS™-10, KISS™-12 .................. 47 Environment................... 48 KISS™...
  • Page 13: Functional Description

    Functional Description General Information The GE Healthcare Electrode Application System KISS™ is used to assist in the application of ECG electrodes for the acquisition of resting and exercise ECGs. The system is optimally suited from an ergonomic viewpoint. A number of different KISS™ versions are available: •...
  • Page 14 Functional Description 22745010H KISS™ / KISS™ Multilead...
  • Page 15: Operating Controls And System Components

    Operating Controls and System Components KISS™ 22745010H KISS™ / KISS™ Multilead...
  • Page 16 Operating Controls and System Components Application System KISS™ and pole Item Description Item Description Electrode distributor Power connector for pump module Negative pressure control Pump module Electrode leads with valve Screw clamp for fixation of pole to table Silver silver chloride electrode Wall-mount facility for pole Filter disk =>...
  • Page 17: Kiss™ Multilead

    Suction pump connector Electrode lead hanger Power cord hanger CAM USB A/T KISS™ interface Electrode distributor Connects to the USB port of the PC Negative pressure control AC power adapter Electrode leads with valve (A1 - A4...
  • Page 18 Operating Controls and System Components 22745010H KISS™ / KISS™ Multilead...
  • Page 19: Testing Instructions For Kiss™ And Kiss™ Multilead

    Testing instructions for KISS™ and KISS™ Multilead Visual Check Check the application system for external intactness: • The distributor housing is undamaged. • The power supply unit with the integrated suction pump is undamaged. • The trunk cable and the electrode leads are intact. No cracks, kinks or deformations.
  • Page 20: Equipment Check For Leaks In The Complete System, Pump, Distributor, Electrode Leads And Suction Cups

    “1. Test Plate (p/n 30344490)” on page 45 • Y-adapter (for connection negative pressure tubing <-> PTT) “2. Y-Adapter (p/n 30344489)” on page 45 • Dummy plug for distributor “4. Dummy Plug for KISS™ (p/n 30344488)” on page 46 22745010H KISS™ / KISS™ Multilead...
  • Page 21: Complete System Test

    Testing instructions for KISS™ and KISS™ Multilead Complete System Test Start up the pump by attaching an electrode. Then remove the electrode. Measure the negative pressure after about 10 seconds: the reading must be between 225 and 265 mmHg (300 and 350 mbar).
  • Page 22: Pump Leakage

    Testing instructions for KISS™ and KISS™ Multilead For pump leaks, proceed as described in “Pump Leakage” on page 22. If the distributor leaks, the entire distributor should be replaced (see Chapter 5 “Spare Parts List (KISS™/KISS™ Multilead)” on page 41).
  • Page 23: Pump Leakage Test With Pumps Requiring Negative Pressure To Start

    Testing instructions for KISS™ and KISS™ Multilead Pump Leakage Test with Pumps Requiring Negative Pressure to Start Start the pump by pulling gently on the syringe piston. Then switch off the negative pressure pump (ECG recorder or power supply unit): The drop in negative pressure must not exceed max.
  • Page 24: Testing Electrical Function (Kiss™)

    See Chapter 5. “Spare Parts List (KISS™/KISS™ Multilead)” on page 41. Testing Electrical Function (KISS™ ™ ™ ) Resistance between the leadwire connector and the KISS™ electrode connector. The measured resistance must be 10 kohm ±1kohm. CAUTION: It is prudent to make this measurement without electrodes as the silver-silver chloride electrodes must not come into contact with other metals.
  • Page 25: Testing Electrical Function (Kiss™ Multilead)

    Testing instructions for KISS™ and KISS™ Multilead Testing Electrical Function (KISS™ Multilead) Resistance between the leadwire connector and the KISS™ electrode connector. The measured resistance must be 10 kohm ±1kohm. CAUTION: It is prudent to make this measurement without electrodes as the silver-silver chloride electrodes must not come into contact with other metals.
  • Page 26: Disassembling The Trunk Cable And The Pcb

    Testing instructions for KISS™ and KISS™ Multilead Disassembling the Trunk Cable and the PCB • First carry out “Disassembling and Assembling the Controller and the End Piece on the Distributor” on page • Disconnect plug from PCB. • Position a pair of tweezers under the end of the tubing and ease the tubing off the extension tube.
  • Page 27: Assembling The Controller

    Testing instructions for KISS™ and KISS™ Multilead Assembling the Controller Slide on controller until all spring catches lock in. Final, check by pulling gently on the controller. KISS™ Multilead - Disassembling, Assembly Disassembling and Assembling the Controller Remove all electrode leads on the distributor. There are 4 spring catches to be opened for each lead.
  • Page 28: Air Tube And Input Pcb

    Testing instructions for KISS™ and KISS™ Multilead • You can remove the connector housings by pulling them carefully off the contact pins. Air Tube and Input PCB The distributor block is available as a spare part (without regulator). The regulator must be ordered separately, if needed.
  • Page 29: Exchange Of The Kiss Multilead Air Tube

    Testing instructions for KISS™ and KISS™ Multilead If this component is to be replaced, a device label must be ordered at techsupport.gemsit@ med.ge.com using the existing Cat. No. and S/N. Exchange of the KISS Multilead Air Tube • Remove all electrode leads from the distributor block and loosen the screws as shown on figure 1 and figure 2.
  • Page 30 Testing instructions for KISS™ and KISS™ Multilead • Pull the locking plates (A) on both sides of the distributor block by using a suitable tool (see figure 4). Figure 4 • Lift the two spring catches on both sides (see figure 5). CAUTION: Do NOT...
  • Page 31 Testing instructions for KISS™ and KISS™ Multilead • Remove the cover from the distributor block (see figure 6). Figure 6 • Pull off the air tube from the distributor block (see figure 7). 22745010H KISS™ / KISS™ Multilead...
  • Page 32 Testing instructions for KISS™ and KISS™ Multilead • Slide the air tube off the cover as shown in figure 8. • Insert the new air tube in the cover as shown in figure 9. Figure 9 22745010H KISS™ / KISS™ Multilead...
  • Page 33 Testing instructions for KISS™ and KISS™ Multilead • Widen the end of the air tube (see A on figure 10) by using a suitable tool so that it fits over the nozzle of the tube connector. Push the air tube over the nozzle. Assemble the cover and the distributor block.
  • Page 34: Kiss™/Kiss™ Multilead O-Ring Replacement

    Testing instructions for KISS™ and KISS™ Multilead KISS™/KISS™ Multilead O-Ring Replacement Remove all electrode leads on the distributor. There are 4 spring catches to be opened. • Lift up two spring catches on one side with a pair of pointed tweezers (Caution: Do no overstretch!), insert pointed tweezers, carefully press down on the spring catch to unlock it and lift component slightly.
  • Page 35 Testing instructions for KISS™ and KISS™ Multilead • Confirm that the O-Rings are correctly positioned. • Perform the following tests: • “Visual Check” on page 19 • “System Function Check” on page 19 • “Electrode Lead Valve Leak” on page 21 •...
  • Page 36 Testing instructions for KISS™ and KISS™ Multilead 22745010H KISS™ / KISS™ Multilead...
  • Page 37: Multilead)

    Use a moist cloth to wipe the pole and the CAM-USB A/T KISS™ interface clean. Do not allow liquid to enter the equipment. When the surface is very soiled, use a mild soap solution or detergent.
  • Page 38: Cleaning Electrodes And Electrode Care

    • Observe the information given in the KISS™ Multilead Operator Manual. For a system in frequent use replace the filter disks once a week! Replacing Leadwires The leadwires are subject to wear and tear and, therefore, they can be replaced.
  • Page 39: Troubleshooting

    Maintenance, Cleaning and Troubleshooting (KISS™/KISS™ Multilead) Troubleshooting Possible error: Electrodes do not become attached by suction, insufficient suction, or electrodes fall off Causes: No or insufficient negative pressure due to: • Kink in negative pressure tubing • Negative pressure tubing connection to pump interrupted •...
  • Page 40 Maintenance, Cleaning and Troubleshooting (KISS™/KISS™ Multilead) 22745010H KISS™ / KISS™ Multilead...
  • Page 41: Spare Parts List (Kiss™/Kiss™ Multilead)

    Spare Parts List (KISS™/KISS™ Multilead) Single Leadwires for KISS™ and KISS™ Multilead 303 443 65 Leadwire C1 303 443 66 Leadwire C2 303 443 67 Leadwire C3 303 443 68 Leadwire C4 303 443 69 Leadwire C5 303 443 70...
  • Page 42 Trunk cable for 10-lead KISS™, universal (non-GE ECG devices) 223 414 09 Trunk cable, 12-lead KISS™, Schwarzer 223 414 10 Trunk cable, 10-lead KISS™ (MicroSmart / MAC™ 1200 / MAC™ 1100) 223 414 15 Trunk cable, 12-lead KISS™ (MicroSmart / MAC™ 1200 / MAC™ 1100) 22745010H KISS™...
  • Page 43 Distributor block for KISS™ Multilead 303 444 27 Blind plug 2056303-001 FRU ASSY TUBE AIR KISS MULTILEAD Miscellaneous Adapters for KISS™ with Universal Connector 223 361 10 for MAC™ 1100 / 1200 and MAC™ 500 (MicroSmart) 223 361 18 for AM4 and AM5...
  • Page 44 Spare Parts List (KISS™/KISS™ Multilead) 22745010H KISS™ / KISS™ Multilead...
  • Page 45: Testing Instructions For Kiss™ And Kiss™ Multilead

    Testing Instructions for KISS™ and KISS™ Multilead 1. Test Plate (p/n 30344490) 2. Y-Adapter (p/n 30344489) 3. Adapter Extension with Syringe (p/n 2081397-001) 22745010H KISS™ / KISS™ Multilead...
  • Page 46: Dummy Plug For Kiss™ (P/N 30344488)

    Testing Instructions for KISS™ and KISS™ Multilead 4. Dummy Plug for KISS™ (p/n 30344488) NOTE: Model 1 was replaced with model 2. 22745010H KISS™ / KISS™ Multilead...
  • Page 47: Specifications

    7 = bridge to 8 (only with 12 electrodes) 8 = bridge to 7 (only with 12 electrodes) 9 = R 10 = L 11 = F 12 = C1 13 = Nst 14 = N 15 = Nax 22745010H KISS™ / KISS™ Multilead...
  • Page 48: Environment

    • atmospheric pressure between 700 and 1060 hPa Storage and transport • KISS w/o pump: temperature between -30°C and +70°C KISS with pump and KISS pump: temperature between -30°C and +60°C • relative humidity between 10 and 95%, non-condensing • atmospheric pressure between 500 and 1060 hPa Weight electrode distributor with 12 electrodes approx.
  • Page 49: Environment

    II; for rated line voltages from 100 to 240 VAC, 50 to 60 Hz (without switchover) 100 to 240 VAC input operating voltage range 700 mA-350 mA input current 12 V output voltage output current 2.5 A 22745010H KISS™ / KISS™ Multilead...
  • Page 50 Specifications Dimensions (pump module) • length 130 mm • width 65 mm • height 70 mm Weight pump module approx. 410 g, incl. power supply: 680 g 22745010H KISS™ / KISS™ Multilead...
  • Page 51 Schematics Assembly Drawings 216 121 01/08 Electrode Application System KISS™ 10 216 122 01 Electrode Application System KISS™ 12 2022865-001 Electrode Application System KISS™ Multilead 22745010H KISS™ / KISS™ Multilead...
  • Page 52 Schematics 22745010H KISS™ / KISS™ Multilead...
  • Page 54 +1 800 558 7044 (US Only) Fax: +49 761 45 43 -233 Shanghai, People's Republic of China 201203 Tel: +86 21 3877 7888 Fax: +86 21 3877 7402 GE Medical Systems Information Technologies, Inc., a General Electric Company, going to market as GE Healthcare. www.gehealthcare.com...