Specifications

KISS™-10, KISS™-12
Electrode application system for acquiring ECG signals from a patient
Electrode distributor
for 10 or 12 plug-in electrode leads
Electrodes
silver-silver chloride electrodes, with connectors
Defibrillation protection
by means of integrated current and voltage limitation
Signal connector
15-pin connection, IEC pin configuration
1 = C2
2 = C3
3 = C4 = Nap
4 = C5
5 = C6
6 = shield
7 = bridge to 8 (only with 12 electrodes)
8 = bridge to 7 (only with 12 electrodes)
9 = R
10 = L
11 = F
12 = C1
13 = Nst
14 = N
15 = Nax