Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The AFFINITY COMPACT is intended for use in the electromagnetic environment specified below. The
customer or the user of the AFFINITY COMPACT should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN
IEC 60601 Test
level
+8 kV contact
+15 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5% UT
(>95% dip in UT) for 0.5
cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5
sec
30 A/m
Compliance
+8 kV contact
+15 kV air
+2 kV for power
supply lines
+1 kV for input/output
lines
+1 kV differential
mode
+2 kV common mode
< 5% UT (>95% dip in
UT)
for 0.5 cycle
40% UT (60% dip in
UT) for
5 cycles
70% UT (30% dip in
UT) for
25 cycles
<5% UT
30 A/m
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the AFFINITY
COMPACT requires continued
operation during power mains
interruptions, it is
recommended that the
AFFINITY COMPACT be
powered from an
uninterruptable power supply
or its battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
environment.
Page 77