5
TENs test requires an additional license for the AC440. If not purchased, the stimulus is grayed out.
D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN
The Output list for channel 2 provides the option to test through head
phones, free field speakers, insert phones or insert masking phone. Note
that the system only shows the calibrated transducers.
The Input list for channel 2 provides the option to select pure tone, warble
tone, narrow band noise (NB), white noise (WN) and TEN noise
Note that the background shading is according to the side that is selected,
red for right, blue for left, and white when off.
Pulsation allows for single and continuous pulsating presentation. The
duration of the stimulus can be adjusted in the AC440 setup.
Sim/Alt allows toggling between Simultaneous and Alternate
presentation. Ch1 and Ch2 will present the stimulus simultaneously when
Sim is selected. When Alt is selected, the stimulus will alternate between
Ch1 and Ch2.
Masking indicates if channel 2 is currently in use as a masking channel and
in that way makes sure masking symbols are used in the audiogram. For
example in paediatric testing through free field speakers, channel 2 can be
set as a second testing channel. Note that a separate store function for
channel 2 is available when channel 2 is not used for masking.
dB HL Increase and Decrease buttons allows for increasing and
decreasing the intensities of channel 1 and 2.
The arrow keys on the PC keyboard can be used for in-/decreasing channel
1 intensities.
PgUp and PgDn on the PC keyboard can be used for in-/decreasing
channel 2 intensities.
Stimuli or attenuator buttons will light up when the mouse goes over and
indicates the active presentation of a stimulus.
A right mouse click in the Stimuli area will store a no response threshold. A
left mouse click in the Stimuli area will store the threshold at the current
position.
Channel 1 stimulation can also be obtained by pressing the space bar or left
Ctrl key on the PC keyboard.
Channel 2 stimulation can also be obtained by pressing the right Ctrl key on
the PC keyboard.
Mouse movements in the Stimuli area for both channel 1 and channel 2 can
ignored depending on the setup.
Frequency and Intensity display area shows what is currently presented.
To the left the dB HL value for channel 1 is shown and to the right for
channel 2 In the centre the frequency is displayed.
Notice that the dB dial setting will flash when trying to go louder than the
maximum available intensity.
Frequency increase/decrease increases and decreases the frequency
respectively. This can also be obtained using the left and right arrow keys
on the PC keyboard.
.