Interacoustics Affinity Compact Instructions For Use Manual page 39

Hide thumbs Also See for Affinity Compact:
Table of Contents

Advertisement

The WR Table
The word recognition (WR) table allows for measuring multiple WR scores using different parameters (e.g.
Transducer, Test Type, Intensity, Masking, and Aided).
Upon changing Transducer, Masking, and/or Aided re-testing an additional WR entry will appear in the WR
table. This allows for multiple WR measurements to be shown in the WR table.
Please refer to the Callisto Additional Information document for more information about Word Recognition
testing.
Binaural and Aided options
To perform binaural speech tests:
1. Click on either SRT or WR, to choose the test to be conducted binaurally
2. Ensure that the transducers are set up for binaural testing. For example, insert Right in channel 1
and insert Left in channel 2
3. Click on
4. Proceed with the test; when stored, results will be stored as binaural results
To perform an aided test:
1. Select the desired transducer. Typically, aided testing is done in the Free Field. However, in certain
conditions, it could be possible to test deeply inserted CIC hearing instruments under headphones,
which would show ear-specific results
2. Click on the Aided button
3. Click on the Binaural button if the test is done in the Free Field so that the results are stored for both
ears at the same time
4. Proceed with test; results will then be stored as aided by showing an Aided icon
D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN
Page 35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ac440Rem440Vsp440 hit440Vsp440Hit440

Table of Contents