Connecting a Mercury Trap (for AA only)

Use a sharp scalpel to cut two suitable lengths of the supplied 6 mm
diameter clear PVC tubing, and connect them to the two cell ports.
Connect the free ends to a 'T'-piece and then to a suitable mercury
trap in accordance with local regulations governing the discharge of
mercury vapor.
Fitting is easier if the ends of the tubing are first softened in hot water.
TIP
Agilent Vapor Generation Accessory VGA 77 and VGA 77P User's Guide
Measure the minimum length of the supplied polyethylene tubing
required to connect the free end of the black fluoro-elastomer
tubing to the sample inlet of the ICP nebulizer. Neatly cut the
required length and make the connection.
, the drain peristaltic pump tubing from the