3. Follow the on-screen prompts.
4. When complete, the Gage will double-beep and display dashed lines.
Material
If a known thickness of the material is not available, but the material is known, this quick
adjustment allows the user to load one of several preprogrammed material velocities.
Cal Reset
Restores the gage back to factory calibration. The factory calibration icon will appear on
the display.
2 Pt Adjust
A 2-Point adjustment allows for greater accuracy while simultaneously adjusting probe
Zero.
Select two reference standards as close as possible in composition to the intended
application. For best results, the thickness of the thicker reference standard should be
equal to, or slightly greater than the thickest part to be measured. The thickness of the
thinner reference standard should be equal to, or slightly less than the thinnest part to
be measured.
1. Measure the thinner reference sample.
2. Lift the probe from the sample and adjust the display down (-) or up (+) to the known
thickness.
3. When the expected thickness is reached, press the center navigation button.
4. Measure the thicker reference sample.
5. Lift the probe from the reference sample and adjust down (-) or up (+) to the known
thickness.
6. Press the center navigation button to store the adjustment and exit.
Cal Lock
Page 5