Statistics

Menu option for configuring Statistics and HiLo Alarm modes.
Statistics
Statistics mode continually displays/updates average, standard deviation, min/max
thickness and number of readings while measuring.
When selected, the statistics icon and statistical summary will appear on the display.
Ignore the last measurement by pressing the (-) button. Press (+) to clear statistics.

Clear

Clears all on-screen Statistics and HiLo tabulations.

HiLo Alarm

Allows the Gage to visibly and audibly alert the user when measurements exceed
user-specified limits.
When HiLo Alarm is selected, the current Lo setting is displayed. Adjust using the (-) or
(+) buttons. Select Next to accept this value. The current Hi setting is now displayed.
Repeat to adjust the Hi setting.
Each measurement will be compared to the defined Hi and Lo limits. The Gage beeps if
results are within those limits. A single low tone will sound if the reading is below the Lo
limit, and a high tone if it is above the Hi limit. Press (+) to clear HiLo readings.
The Statistics icon will appear on the display.
Page 3