DeFelsko PosiTector Series Calibration Procedure

DeFelsko PosiTector Series Calibration Procedure

Ultrasonic thickness gage

Advertisement

Quick Links

PosiTector UTG, UTG-STD, UTG-ME, UTG-
C, UTG-CA, UTG-CLF, UTG-CX, UTG-M &
Introduction and UUC Performance Requirements ............................................................................................ 2 
Table 1-1 Measurement Ranges ......................................................................................................................... 2 
Measurement Standards and Support Equipment Performance Requirements .................................................. 2 
Table 2-1 UUC Accuracy Requirements and Description ................................................................................. 2 
Table 2-2 Minimum Use Specification .............................................................................................................. 3 
Table 2-3 Actual Equipment Specification ........................................................................................................ 3 
Table 2-4 Calibration Environmental and Warm Up Requirements .................................................................. 3 
Preliminary Operations ....................................................................................................................................... 3 
Calibration Process ............................................................................................................................................. 4 
Performance Requirements ................................................................................................................................ 5 
Table 5-1 Performance Requirements and Calibration Data for PosiTector UTG ............................................. 5 
Management Procedure Change Notice ........................................................................................................................ 6 
Management Procedure 2585 Rev. F
Calibration Procedure
DeFelsko Corporation
UTG-P
Ultrasonic Thickness Gage

Table of Contents

Management Procedure 2585
Revision: F
Date Issued: Feb 21, 2003
Date Revised: June 10, 2019
1

Advertisement

loading

Summary of Contents for DeFelsko PosiTector Series

  • Page 1: Table Of Contents

    Management Procedure 2585 Revision: F Date Issued: Feb 21, 2003 Date Revised: June 10, 2019 Calibration Procedure DeFelsko Corporation PosiTector UTG, UTG-STD, UTG-ME, UTG- C, UTG-CA, UTG-CLF, UTG-CX, UTG-M & UTG-P Ultrasonic Thickness Gage Table of Contents 1  Introduction and UUC Performance Requirements .................... 2 ...
  • Page 2: Introduction And Uuc Performance Requirements

    Introduction and UUC Performance Requirements This procedure describes the calibration of the PosiTector UTG Ultrasonic Thickness Gage and probe. Table 1-1 Measurement Ranges Models Measurement Range * Velocity Range UTG, UTG-STD, UTG-C 1.0 - 125 mm UTG-CA, UTG-CLF, UTG-CX (0.040 - 5.000”) 1,250 to 10,000 m/s 2.5 - 125 mm UTG-M &...
  • Page 3: Preliminary Operations

    Table 2-2 Minimum Use Specification Model Range Accuracy UTG, UTG-STD, UTG-C 1.0 - 125 mm ± 0.008 mm UTG-CA, UTG-CLF, UTG-CX (0.040 - 5.000”) (± 0.00025”) 2.5 - 125 mm ± 0.008 mm UTG-M & UTG-ME (0.100 - 5.000”) (± 0.00025”) 0.200 –...
  • Page 4: Calibration Process

    Note: Please refer to UUC instruction manual for menu navigation instructions, details on features and operating instructions (i.e. required amount of couplant and pressure). 3.2.1 For bodies with serial number after 700,000; when the unit is powered down, simultaneously hold the “+” and middle buttons until the rest symbol appears. All other bodies press and hold the “+”.
  • Page 5: Performance Requirements

    Performance Requirements Note: The technician will collect the data needed to complete column B of Table 5-1. Do not write in this procedure. Table 5-1 Performance Requirements and Calibration Data for PosiTector UTG Reference Min. Reading Max. Reading Actual Gage Thickness Allowed Allowed...
  • Page 6: Management Procedure Change Notice

    Management Procedure Change Notice Procedure Number: MP 2585 Revision Level: F Date of Change: June 10 , 2019 Title: Calibration Procedure for PosiTector UTG, UTG STD, UTG-C, UTG-CA, UTG-CLF, UTG-CX, UTG-M, UTG- ME & UTG-P Reason for Change: • New products Description of Change: •...