The Midwest
Handpiece Maintenance Instructions
Refer to DFU included with handpiece for detailed instructions on operation,
sterilization, warranty and ordering Information. Additional information can be
obtained at www.stylusatc.com or by calling 1.800.800.7202
Weekly
1
O-Rings
2
As Needed
1
2
www.StylusATC.com
©2010 DENTSPLY. All rights reserved.
Stylus
ATC
®
Lubricate chuck once a week
a. Weekly, or if push button gets stiff,
insert a drop of Midwest Plus
lubricant (REF380130) into
handpiece chuck only. Do this
before sterilization.
b. Free run handpiece without a bur
until excess lubricant is expelled
(minimum 30 seconds).
Wipe off excess lubricant on
outside of handpiece with dry
gauze pad or paper towel.
Lubricate coupler O-Rings
Weekly or if handpiece swivel gets stiff,
lightly lubricate the coupler O-Rings with
Midwest Plus
lubricant (REF380130).
Clean handpiece water spray holes
Run the cleaning wire back and forth in
spray holes to loosen and remove
any mineral build-up at spray surface.
Contents of the Maintenance Kit
1
2
3
4
1 Cleaning Wires
2 Maintenance Coupler
(880040)
3 Bulbs
(880070)
4 O-Rings
5 Lubrication Nozzle and Cap
(380150)
6 Spray Guard
7 Midwest Plus
Technical Customer Service / Air Repair