Operation

4294A Setting

4294A Setting
Before you begin your measurement, you should perform the adapter setup in the
4294A. Also refer to the operation manual of the 4294A about the adapter setup.

Adapter Setup

Connect the 16048G/H to the 4294A and perform the adapter setup described
below.
For adapter setup, use the 100 Ω Resistor furnished with the 4294A (Agilent P/N
NOTE
04294-61001).
1.
2.
3. Press
4. Select
5.
6. Connect the Lcur terminal and Lpot terminal on the 16048G or 16048H to the
Connecting 100 Ω Resister (Phase Compensation Data Measurement)
Figure 3-1
18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
More than 30 minutes warm-up time is required after turning on the 4294A.
Press [Cal] key to bring up the Calibration Menu.
key to bring up the adapter setup Menu. [ ] shows current
ADAPTER [ ]
settings.
in the case of 16048G,
4TP 1M
selection is completed, the softkey label will be underlined.
Press SETUP key to bring up the Adapter Setup Menu.
Hpot terminal and Hcur terminal of the 100 Ω Resistor furnished with the
4294A. The Hcur and Hpot terminal on the 16048G/H should remain open
(nothing should be connected).
in the 16048H. When the
4TP 2M
Chapter 3