Agilent Technologies 16191A Operation And Service Manual

Side electrode smd test fixture

Advertisement

Quick Links

A l l t e s t I n s t r u me n t s , I n c .
5 0 0 C e n t r a l A v e .
F a r mi n g d a l e , N J 0 7 7 2 7
P : ( 7 3 2 ) 9 1 9 - 3 3 3 9
F : ( 7 3 2 ) 9 1 9 - 3 3 3 2
a l l t e s t . n e t
s s a l e s @ a l l t e s t . n e t
T h e t e s t & me a s u r e me n t
e q u i p me n t y o u n e e d a t
t h e p r i c e y o u w a n t .
A l l t e s t c a r r i e s t h e w o r l d ' s l a r g e s t s e l e c t i o n o f
u s e d / r e f u r b i s h e d b e n c h t o p t e s t & me a s u r e me n t
e q u i p me n t a t 5 0 % t h e p r i c e o f n e w .
O O u r e q u i p me n t i s g u a r a n t e e d w o r k i n g , w a r r a n t i e d , a n d
a v a i l a b l e w i t h c e r t i f i e d c a l i b r a t i o n f r o m o u r i n - h o u s e s t a f f
o f t e c h n i c i a n s a n d e n g i n e e r s .
• 1 0 + f u l l t i me t e c h n i c i a n s w i t h o v e r 1 5 0 y e a r s o f
s p e c i a l i z a t i o n
• 9 0 d a y w a r r a n t y & 5 d a y r i g h t o f r e t u r n o n a l l
e q u i p me n t
• • 1 - 3 y e a r w a r r a n t i e s f o r n e w a n d
p r e mi u m- r e f u r b i s h e d e q u i p me n t
• E v e r y u n i t t e s t e d t o O E M s p e c i f i c a t i o n s
• S a t i s f a c t i o n g u a r a n t e e d
Y o u h a v e p l a n s , w e w i l l h e l p y o u a c h i e v e t h e m.
A n y p r o j e c t . A n y b u d g e t .
t
G e t a q u o t e t o d a y !
C C a l l ( 7 3 2 ) 9 1 9 - 3 3 3 9 o r e ma i l s a l e s @a l l t e s t . n e t .

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 16191A

  • Page 1 T h e t e s t & me a s u r e me n t e q u i p me n t y o u n e e d a t t h e p r i c e y o u w a n t . A l l t e s t I n s t r u me n t s , I n c .
  • Page 2 Agilent 16191A Side Electrode SMD Test Fixture Operation and Service Manual Agilent Part No. 16191-90040 Printed in JAPAN May 2003 Sixth Edition...
  • Page 3 No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd. Component Test PGU-Kobe 1-3-2, Murotani, Nishi-ku, Kobe-shi, Hyogo, 651-2241 Japan Copyright Agilent Technologies Japan, Ltd. 1993, 1996, 1999, 2000, 2001, 2003...
  • Page 4 Manual Printing History The manual printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The manual part number changes when extensive technical changes are incorporated.
  • Page 5 WARNINGS given elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer's failure to comply with these requirements. DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of ammable gasses or fumes.
  • Page 6 Instrument Speci cations of this manual, the warranty shall be for the speci ed period. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective.
  • Page 7 No other warranty is expressed or implied. Agilent Technologies speci cally disclaims the implied warranties of merchantability and tness for a particular purpose. Exclusive Remedies The remedies provided herein are buyer's sole and exclusive remedies.
  • Page 8 Safety Symbols General de nitions of safety symbols used on equipment or in manuals. Instruction manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument.
  • Page 9 APC-7 is a U.S. registered trademark of the Bunker Ramo Corporation. viii...
  • Page 10: Table Of Contents

    Contents 1. General Information Introduction ....... . . Manual Summary ......Product Description .
  • Page 11 1-3. Option 010 Shorting Device Residual Inductance (Typical) ..2-1. 16191A Contents ......
  • Page 12: General Information Introduction

    General Information Introduction The purpose of this manual is to enable you to use your 16191A Side Electrode SMD Test Fixture eciently and con dently. Manual Summary This manual contains the following: The speci cations of the 16191A (see this chapter). Installing the 16191A (see Chapter 2).
  • Page 13: Options Available

    Options Available The following option is supplied for the 16191A: Option 010 Add industry standard size short bar set Option 701 Short bars set (1x1x2.4, 1.6x2.4x2, 3.2x2.4x2.4, 4.5x2.4x2.4)mm Option 710 Add the magnifying lens and tweezers Accessories Supplied The following accessories are supplied with the 16191A: Table 1-1.
  • Page 14: Speci Cations

    Speci cations This section lists the complete 16191A speci cations. These speci cations are the performance standards and limits against which the 16191A is tested. When shipped from the factory, the 16191A meets the following speci cations: Maximum DC Bias Voltage : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :...
  • Page 15: Residual Inductance Of The Shorting Bar

    Figure 1-1. Applicable DUT Size Residual Inductance of the Shorting Bar The usual method to compensate the test xture's residual inductance is to let SHORT = 0H. In this method, the measurement result is the relative value of the measured impedance to the shorting bar's impedance.
  • Page 16 Table 1-3. Option 010 Shorting Device Residual Inductance (Typical) Shorting Bar l [mm] d [mm] h [mm] O set [mm] Connector Inductance (Typical) P/N 16191-29005 0.75 3.5 mm 0.5 nH P/N 16191-29006 0.45 3.5 mm 0.4 nH P/N 16191-29007 7 mm 0.9 nH P/N 16191-29008 7 mm...
  • Page 18: Preparation For Use

    Preparation for Use Introduction This chapter explains how to install the 16191A. The topics include the following: Initial inspection. Connecting the test xture for use. Repackaging the test xture for shipment. Measurement error correcting function for each instrument. Initial Inspection The side electrode SMD test xture has been carefully inspected before being shipped from the factory.
  • Page 19: 16191A Product Overview

    Figure 2-1. 16191A Product Overview Table 2-1. 16191A Contents Description Agilent Part Quantity Number Test xture 16191A Operation and Service Manual P/N 16191-90020 Shorting device 2.4[mm]) P/N 16191-29001 (1.6 2[mm]) P/N 16191-29002 (2.4 3.2[mm]) P/N 16191-29003 (2.4 4.5[mm]) P/N 16191-29004...
  • Page 20: Connecting The Test Fixture For Use

    Connecting the Test Fixture for Use Note Some instruments require an adapter to be connected to the 16191A (see \Applicable Instruments" in Chapter 1). 1. Calibrate your analyzer at the APC-7 connector plane before connecting test xture. See Table 2-2 for the applicable error correcting function. After calibration, disconnect any standards from the APC-7 connector.
  • Page 21: Repackaging The Test Fixture For Shipment

    SMD test xture in heavy paper and pack in anti-static plastic packing material. Use sucient shock absorbing material on all sides of the 16191A to provide a thick, rm cushion and to prevent movement. Seal the shipping container securely and mark it FRAGILE .
  • Page 22: Operation Introduction

    DUT: indicates procedures in the following pages.) Described in Chapter 2. The 16191A's electrical length is 14.0 mm (typical value). To set the electrical length, see your analyzer's Operation Manual. To set the xture and devices for xture compensation, see: \How to set the SHORT condition, and the OPEN condition"...
  • Page 23: Overview

    Overview Figure 3-1. 16191A Test Fixture Overview Electrodes These are connected to the device. (Pressure) arm Fixes the device. Pressure adjustment Adjusts the pressure applied to the device. Arm latch Latches the arm o to turn up and down the arm.
  • Page 24: How To Set The Short Condition, And The Open Condition

    How to set the SHORT condition, and the OPEN condition Setting the SHORT condition Operation...
  • Page 25 Operation...
  • Page 26: Setting The Open Condition

    Setting the OPEN condition Operation...
  • Page 27: How To Connect The Dut

    How to connect the DUT Operation...
  • Page 28: Adjusting The Pressure Arm Stroke

    Adjusting the Pressure Arm Stroke Depending on the device size, change the pressure arm stroke, so that the arm properly presses the device. 1. Loosen the screws with the wrench. 2. Adjust the arm stroke. 3. Tighten the screws. Figure 3-2. Arm Stroke Adjustment Operation...
  • Page 30: Service Introduction

    When you repair the 16191A, put on lint-free groves to avoid contaminating inner parts of the 16191A. Agilent Technologies supplies lint-free groves (Agilent part number 9300-0163). General Information Table 4-1 shows all items included with the 16191A Side Electrode SMD Test Fixture. Service...
  • Page 31: 16191A Product Overview

    Table 4-1. 16191A Product Overview Reference Agilent Part Designator Number Qty. Description (not assigned) Main Assembly 16193-60002 Magnifying Glass 8710-1181 Wrench 8710-2081 Tweezers 16191-29001 Shorting Device (1 2.4 mm) 16191-29002 Shorting Device (1.6 2 mm) 16191-29003 Shorting Device (2.4 3.2 mm) 16191-29004 Shorting Device (2.4...
  • Page 32: Assembly Replacement

    Assembly Replacement This section includes Disassembly and Assembly Procedures and Replacement Parts for the Main Assembly. Disassembly and Assembly Procedure This section contains the information required to disassemble and assemble the Main Assembly. Test Stage, Pressure Arm, and Locking Arm Assembly Removal: 1.
  • Page 33: Replaceable Parts

    Replaceable Parts Table 4-2 through Table 4-6 show and list the replaceable parts for the 16191A. The parts listed can be ordered from your nearest Agilent Technologies Oce. Ordering information must include the Agilent part number and the quantity required.
  • Page 34: Replaceable Parts For Main Assembly (Without Top Cover)

    Table 4-3. Replaceable Parts for Main Assembly (without Top Cover) Reference Agilent Part Designator Number Qty. Description 0515-1550 Screw Pan Head M3L8 16191-04003 Bottom Cover 16191-23001 Shaft (not assigned) Pressure Arm Assembly 16191-23003 Shaft 16191-25005 Bushing 16191-23010 Gear 16191-24017 Screw 16191-24016 Horizontal Positioning Knob 16191-04002...
  • Page 35: Replaceable Parts For Locking Arm Assembly

    Table 4-4. Replaceable Parts for Locking Arm Assembly Reference Agilent Part Designator Number Qty. Description 16191-24005 Knob 16191-24019 Holder 0510-0015 E-ring Service...
  • Page 36: Replaceable Parts For Pressure Arm Assembly

    Table 4-5. Replaceable Parts for Pressure Arm Assembly Reference Agilent Part Designator Number Qty. Description 0515-1050 Screw Socket Head M3L8 0510-0976 Screw Pan Head M2L6 16191-25003 Holder 16191-24018 Block 16191-29011 Spring 16191-25006 Washer 16191-01203 Angle 16191-00602 Plate 16191-24006 Screw 16191-24010 PPNR007783 Spring 16191-23005...
  • Page 37: Replaceable Parts For Test Stage

    Figure 4-1. Replaceable Parts for Test Stage Service...
  • Page 38: Replaceable Parts For Test Stage Assembly

    Table 4-6. Replaceable Parts for Test Stage Assembly Reference Agilent Part Designator Number Qty. Description 0515-0914 Screw Flat Head M3L6 16191-00201 Table 16191-24022 Screw 16191-24014 Guide 16191-25004 Guide 16191-24015 Contact-L 0510-0045 E-Ring 16191-24021 0510-0952 Screw Flat Head M2L4 16191-00601 Ground Plate 0515-2421 Screw Pan Head M1.4L3 16191-25002...
  • Page 39: Troubleshooting

    Troubleshooting This section includes the functional check procedure used to troubleshoot the 16191A Side Electrode SMD Test Fixture. Open Impedance Check 1. Adjust the electrode distance for connecting the 1.6 2.4 mm Shorting Device (16191-29002). 2. Place the xture on the calibrated APC-7 terminal of the 4291B RF Impedance Analyzer.
  • Page 40: Short Impedance Repeatability Check

    Table 4-8. Short Impedance Value Guideline Parameter Frequency Guideline (absolute value) 100 MHz 3 nH 1 nH 1 GHz 3 nH 1 nH Short Impedance Repeatability Check 1. After the Short Impedance Check is completed, remove the Shorting Device and put it back on the same place again.
  • Page 42 REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.