Agilent Technologies 16043A Operation And Service Manual
Agilent Technologies 16043A Operation And Service Manual

Agilent Technologies 16043A Operation And Service Manual

3-terminal smd test fixture
Table of Contents

Advertisement

Quick Links

Agilent 16043A/B 3-terminal SMD Test Fixture
Operation and Service Manual
Second Edition
Agilent Part No. 16043-90010
May 2000
Printed in Japan

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 16043A

  • Page 1 Agilent 16043A/B 3-terminal SMD Test Fixture Operation and Service Manual Second Edition Agilent Part No. 16043-90010 May 2000 Printed in Japan...
  • Page 2 WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. •...
  • Page 3 Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
  • Page 4 Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility, or to the...
  • Page 5 IMPORTANT No other warranty is expressed or implied. Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose. Exclusive Remedies The remedies provided herein are buyer’s sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.
  • Page 7: Table Of Contents

    Connecting the 16043A/B to a Measuring Instrument ........
  • Page 8 Adjusting Ball Plunger (16043A only) ........
  • Page 9: Installation Guide

    Installation Guide This chapter describes the procedures required from when you receive the 16043A/B Test Fixture until you can start using it.
  • Page 10: Incoming Inspection

    Agilent Technologies Sales office. If the shipping container is damaged or the cushioning material shows signs of unusual stress, notify the carrier as well as the Agilent Technologies office. Keep the shipping materials for the carrier's inspection.
  • Page 11 Installation Guide Incoming Inspection Figure 1-1 16043A Package Contents Chapter 1...
  • Page 12 Installation Guide Incoming Inspection Table 1-2 16043B Package Contents Name Agilent Part Number Qty. 16043B 3-terminal SMD Test Fixture 16043-01203 Support Angle for the 4294A 16043-00601 Device Guide 0515-2791 Screws for the Device Guide 16043-24004 Guide Pin Support Angle 16043-01212 Tweezer 8710-2081 Positioning Tool...
  • Page 13 Installation Guide Incoming Inspection 16043B Package Contents Figure 1-2 Chapter 1...
  • Page 14: Connecting The 16043A/B To A Measuring Instrument

    For details, see “Fabricating the Device Guide” on page 27 and “Mounting the Device Guide” on page 32. Connectable Measuring Instruments The 16043A/B Test Fixture can be connected to a Measuring instrument using four BNC connectors. Table 1-3...
  • Page 15: Connecting To 4294A

    4294A and gently press the fixture straight in as long as it will go. Step 2. Turn the two BNC connector rotation levers of the 16043A/B to align the positions of the connector grooves. Then press straight in until the two units come into complete contact.
  • Page 16: Connecting To Other Instruments Than 4294A

    Figure 1-4 Support Angle Step 1. Turn the two BNC connector rotation levers of the 16043A/B to align the positions of the connector grooves, and align the connectors with the UNKNOWN terminal. Then gently press straight in until the two units come into complete contact.
  • Page 17: Product Overview

    Product Overview...
  • Page 18: Product Overview

    Product Overview Product Overview 16043A and 16043B are 3-terminal SMD test fixtures. They enable SMD type devices with three terminals to be measured with high precision and measurement repeatability. By replacing the contact board the 16043A/B allows measurement of SMD devices ranging...
  • Page 19: Measurement Principle

    Guard Measurement Using Impedance Instrument Electrode Construction A DUT with electrodes attached on the underside is placed on the fastening electrodes on the contact board of the 16043A/B and pressed down from above to connect. Figure 2-3 Electrode Construction Chapter 2...
  • Page 20 Product Overview Product Overview Figure 2-4 Connection Arrangement Chapter 2...
  • Page 21: Names Of Parts And Functions

    Product Overview Product Overview Names of Parts and Functions The names of each part of 16043A/B are shown in Figure 2-5, and their functions are described in Table 2-1. Figure 2-5 Names of Parts Table 2-1 Names of Parts and Functions...
  • Page 22: Names Of Accessories And Functions

    Product Overview Product Overview Names of Accessories and Functions The names of accessories are shown in Figure 2-6, and their functions are described in Table 2-2. Figure 2-6 Accessories Table 2-2 Names of Accessories and Functions Name Function Guide Pin Used for aligning the position when the device guide is attached.
  • Page 23: Screws For The Device Guide *3 0515-2791

    To enable measurement of DUTs of different sizes, three types of contact boards with different distances between the electrodes are available as options. Contact boards are available for use with 16043A and 16043B. To distinguish the contact board types, figures are imprinted on the boards.
  • Page 24: Additional Device Guide 16043-00601

    Product Overview Product Overview Figure 2-9 Distances Between Electrodes on Contact Boards Table 2-3 Contact Boards and Compatible Chip Sizes Contact Board Compatible Chip Size No.1 8133 - 3731 No.2 3731 - 2520 No.3 2520 - 2012 Chapter 2...
  • Page 25: Operation

    Operation This chapter describes preparations, fixture preparations, and fixture compensation when using the 16043A/B to take measurements as well as DUT connection and measurement methods.
  • Page 26: Flow Of Measurements

    Operation Flow of Measurements Flow of Measurements Follow the steps below when taking DUT measurements with the 16043A/B. 1. Preparation for Measurement Create a device guide and short bar to fit the shape of the DUT to be measured. Also create a short bar that matches the size and shape of the DUT to be measured.
  • Page 27: Preparations For Measurement

    Operation Preparations for Measurement Preparations for Measurement Fabricating the Device Guide The device hole in the device guide determines the position when the DUT or short bar is placed on the electrodes of the contact board. Consequently, measuring with a device hole that does not match the shape of the DUT will prevent proper contact between the DUT and the electrodes and correct measurement will not be accomplished.
  • Page 28: Creating The Short Bar

    Preparations for Measurement Creating the Short Bar Short bars are not furnished with the 16043A/B. Create a short bar that matches the size and shape of the DUT to be measured. The short bar is used to compensate for the same amount of residual short as when the DUT is measured.
  • Page 29: Selecting And Mounting The Contact Board

    To accommodate various sizes and shapes of DUTs, three types of contact boards are optionally available for 16043A and 16043B. While referring to Chapter 2 “Contact Board” on page 23, select a contact board that fits the DUT to be measured. The contact board is mounted in different ways for the 16043A and 16043B.
  • Page 30: Mounting The 16043B Contact Board

    Operation Selecting and Mounting the Contact Board Mounting the 16043B Contact Board Step 1. Remove the screws from the two positions on the upper part of the fixture and remove the device holder block. Figure 3-4 Remove the Device Holder Bolck Step 2.
  • Page 31 Operation Selecting and Mounting the Contact Board CAUTION Do not touch the electrodes with hands or damage them in any way as this could adversely affect the measuring precision and repeatability. Step 3. Tighten the screws to secure the contact board. Step 4.
  • Page 32: Mounting The Device Guide

    Step 1. In the case of the 16043A, move the sliding mechanism all the way back. Step 2. Place the device guide on the contact board and insert the guide pin into the guide hole to align the position.
  • Page 33: Carrying Out Fixture Compensation

    Also, when the 16043A is to be used for load capacity measurement, fixture compensation is necessary to ensure more accurate measurements. For the electrode used for load capacity measurement, take measurements of the data for open compensation and of the data for short compensation.
  • Page 34: Measuring Short Compensation Data

    For normal guard measurement using the 16043A/B, short the electrode's H terminal and the L terminal on the contact board. When measuring load capacity using the 16043A, short the electrode connected to the H terminal and the L terminal on the contact board and make the insulator part contact the G terminal as shown in Figure 3-9.
  • Page 35 Operation Carrying Out Fixture Compensation Figure 3-9 Shorting the Electrodes Using the Short Bar (16043A) Chapter 3...
  • Page 36: Connecting And Measuring Dut

    Operation Connecting and Measuring DUT Connecting and Measuring DUT Connect the DUT to the electrodes and take measurements. Step 1. Using tweezer, align the DUT with the device guide and place on the electrodes. Figure 3-10 Connecting a DUT Step 2. Lower the device holder and turn it counterclockwise to lock and hold the DUT in place. NOTE The force with which the DUT is secured can be adjusted using the device holder pressure regulating screw.
  • Page 37: Measuring A 3-Terminal Device

    Operation Connecting and Measuring DUT Measuring a 3-terminal Device The 16043A enables measurement of all the constituent elements of a 3-terminal device. Figure 3-11 3-terminal Device To measure all the constituent elements of a 3-terminal device, use the sliding mechanism to move the measurement stage to change the electrode of the contact board to which the DUT is connected.
  • Page 38 Operation Connecting and Measuring DUT Figure 3-12 Connection of the Electrodes on the Contact Board of the 16043A The following measurements can be taken by each electrode. 1. Measurement of the characteristics of Z only (1 in Figure 3-12) 2. Measurement of the characteristics of Z only (2 in Figure 3-12) 3.
  • Page 39: Specifications And Supplemental Performance Characteristics

    Specifications and Supplemental Performance Characteristics This chapter describes the specifications and reference data for 16043A/B.
  • Page 40: Specifications

    ≤ 90 %RH (wet-bulb temperature < 65°C) Relative humidity 75 (D) × 105 (W) × 95 (H) mm Dimensions Mass Approximately 500 g (16043A), Approximately 330 g (16043B) Safety standard compliance EN61010-1:1993 +A2:1995 IEC61010-1:1990 +A1:1992 +A2:1995 CSA C22.2 No.1010.1:1992 INSTALLATION CATEGORY I...
  • Page 41: Supplemental Performance Characteristics

    Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics Supplemental Performance Characteristics Useful data for use of the 16043A/B are found in this section. These reference data are not guaranteed as specifications. Additional Errors (When Open/Short Compensation is Conducted) The additional errors when the 16043A/B is connected by two terminals can be found in the following ways.
  • Page 42 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics Figure 4-1 Impedance Additional Error For D Measurement The loss factor additional error De can be found using the impedance additional error Ze [%] as shown in the following. ≤ When Dx De = Ze / 100 ≤...
  • Page 43 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics f: Measurement frequency Csx: Measured value for Cs Rsx: Measured value for Rs Changes in Proportional Errors in Measurement of 3-terminal Device When measuring impedance, the relations between the residual impedance held by the fixture's connectors and the contact resistance and Z (Figure 4-2) have the effect that the actual measured voltage differs from the voltage applied to the device.
  • Page 44: Changes In Measurement Data Caused By Parallel Impedance

    Z and Z The residual amounts of R and L of the guard of the 16043A/B are as follows (reference values). 6 nH or less 20 mΩ or less The measurement waveform changes in accordance with the relationship between R and Z , but in that event it is possible to regard this as described on the following page.
  • Page 45 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics Figure 4-4 Circuit Model When Measuring 3-terminal Device (A) Circuit model when measuring (B) Circuit model for Y-∆ 3-terminal device conversion As shown in Figure 4-4 (A), changes in measurement results occur as parallel impedance is formed in relation to Z by the impedance Z caused by the internal impedance Z...
  • Page 46 Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics The measurement value Z can be expressed as meas < < ------------------- < O G C U < < and Z is found by the following equation. < < -------------------------- - O G C U <...
  • Page 47: Maintenance And Service

    Maintenance and Service This chapter mainly describes cleaning and replacement parts for the fixture.
  • Page 48: Maintenance

    Maintenance and Service Maintenance Maintenance The following describes the maintenance required in order to ensure accomplishment of accurate measurements. Maintenance comprises the following two items. • Cleaning • Replacement of contact board Cleaning To maintain accurate measurements, the electrode sections of the contact board should be cleaned periodically.
  • Page 49: Maintenance Kit

    The 16043U/V maintenance kit is available to provide consumable products and replacement parts for the 16043A/B. 16043U/V Maintenance Kit The 16043U maintenance kit is available for the 16043A, and the 16043V maintenance kit is available for the 16043B. Options are separately available for each kit. Table 5-1...
  • Page 50: Replaceable Parts

    To order parts, specify the Agilent part number and the check digit (C/D). If the part, which is causing problems, is a part that cannot be disassembled, please order the part unit, which the affected part is a part of. Sales and Service Offices of Agilent Technologies also accept products for repairs.
  • Page 51 Maintenance and Service Replaceable Parts Table 5-3 Replaceable Parts (Common for the 16043A/B) Agilent Part No. Description 16047-60015 BNC Connector Assembly 16047-60016 BNC Connector 16380-24001 Terminal 0515-0977 Screw M2L4 Cover 0515-0999 Screw M2.5L4 0515-0914 Screw M3L6 16043-24003 Clevis 16043-01202 Shield...
  • Page 52: Dut Holder

    Maintenance and Service Replaceable Parts DUT Holder Figure 5-3 Exploded View (DUT Holder Common for 16043A/B) Chapter 5...
  • Page 53 Maintenance and Service Replaceable Parts Table 5-4 Replaceable Parts (DUT Holder Common for 16043A/B) Agilent Part No. Description 16043-24013 Screw 16043-24012 Cover 16043-24015 Screw 1460-2636 Spring 16043-25001 1460-2635 Spring 16043-24011 Cylinder 16043-24014 Screw Chapter 5...
  • Page 54: Device Holding Block

    Maintenance and Service Replaceable Parts Device Holding Block Figure 5-4 Exploded View (Device Holding Block for 16043A) Chapter 5...
  • Page 55 Maintenance and Service Replaceable Parts Table 5-5 Replaceable Parts (Device Holding Block 16043A) Agilent Part No. Description 16043-01201 Angle 0515-0924 Screw M3L6 0515-1508 Screw M3L4 0515-2791 Screw M2L3 16043-00601 DUT Guide 16043-20001 Base 1494-0078 Slide Unit 16043-66501 Contact Board No.1(16043A) 16043-66502 Contact Board No.2(16043A)
  • Page 56 Maintenance and Service Replaceable Parts Figure 5-5 Replaceable Parts (Device Holding Block 16043B) Chapter 5...
  • Page 57: Other Parts

    16043A/B. Refer to “Maintenance Kit” on page 49 for detail. Other Parts Table 5-7 Replaceable Parts (Other Parts) Agilent Part No. Description 16043-60150 Carrying Case (for 16043A) 16043-60250 Carrying Case (for 16043B) 16043-24004 8710-2081 Tweezers 16043-00607 Fixture for Position Adjustment...
  • Page 58: Replacement Of Parts

    Maintenance and Service Replacement of Parts Replacement of Parts The following explains the method for replacing the sliding mechanism. Sliding Function (16043A Only) Required Tools • Posidriv screwdriver #2 • Hex wrench 2.5 mm Procedure Step 1. From the underside of the fixture, remove the screws (Figure 5-6 (1)) securing the sliding unit.
  • Page 59 Step 6. Check that the sliding mechanism moves smoothly, and also that it can be fixed at designated positions. If the mechanism moves abnormally, adjust the position of the ball plunger while referring to “Adjusting Ball Plunger (16043A only).” Chapter 5...
  • Page 60: Adjusting Ball Plunger (16043A Only)

    Step 7. From the underside of the fixture, insert screws and secure the sliding mechanism, then remove the pin. Step 8. Mount the specimen holder while referring to “Adjusting Device Holder Position” on page 61. Adjusting Ball Plunger (16043A only) Required Tools • Posidriv screwdriver #2 •...
  • Page 61: Adjusting Device Holder Position

    Step 5. Wait until the LOCTITE hardens. The ball plunger is now held in place. Adjusting Device Holder Position The following explains the procedure for position adjustment of the specimen holder for 16043A. For the procedure for 16043B, refer to “Mounting the 16043B Contact Board” on page 30. Required Tools •...
  • Page 62 Maintenance and Service Replacement of Parts Figure 5-10 Adjusting Device Holder Position Step 4. Secure the cover with screws (Figure 5-10 (4)). Step 5. Raise the specimen holder, and remove the positioning tool. Chapter 5...
  • Page 63: Assembling Check

    Step 3. Use a tester or a multimeter to check continuity. The continuity of each position is shown in Table 5-8. NOTE For the test points, refer to Figure 5-11. Table 5-8 Assembling Check Items 16043A 16043B Path Continuity Path Continuity...
  • Page 64 Maintenance and Service Assembling Check Figure 5-11 Assembling Check Points Chapter 5...