Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
The Concentrator is intended for use in the electromagnetic environment specified below. The user of the Concentrator
should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3V/m
Electrostatic
± 6 kV contact
discharge (ESD)
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power
transient/burst
supply lines
± 1 kV for input/
EC 61000-4-4
output lines
Surge
± 1 kV line(s) to
line(s)
IEC 61000-4-5
± 2 kV line(s) to
earth
Voltage dips, short
<5% U
interruptions and
U
voltage variations
on power supply
40% U
input lines
U
5 cycles
IEC 61000-4-11
70% U
U
<5% U
U
Power frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
96-06947-00-01AG3 User Manual Inogen One G3HF-Multilanguage.indb 41
Compliance
Level
3 Vrms
150 kHz
to 80 MHz
3V/m
80 MHz
to 2.5 GHz
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
(>95% dip in
<5% U
(>95% dip
T
T
) for 0.5 cycle
in U
) for 0.5 cycle
T
T
(60% dip in
40% U
(60% dip
T
T
) for
in U
) for
T
T
5 cycles
(30% dip in
70% U
(30% dip in
T
T
) for 25 cycles
U
) for 25 cycles
T
T
(>95% dip in
<5% U
(>95% dip
T
T
) for 5 sec
in U
) for 5 sec
T
T
3 A/m
3 A/m
Electromagnetic
Environment - Guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the device, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d=1.2√
150 kHz to 80 MHz
P
d=1.2√
80 MHz to 800 MHz
P
d=2.3√
800 MHz to 2.5 GHz
P
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
, should be less than the compliance level in each
a
frequency range
.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the [ME EQUIPMENT or ME SYSTEM] requires
continued operation during power mains interruptions, it is recommended
that the [ME EQUIPMENT or ME SYSTEM] be powered from
an uninterrupted power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical hospital or home environment.
41