Inogen One® G3 Column Change Procedure (continued)
11. Insert column (metal tube) into the Inogen One® G3
concentrator. Do not leave the column exposed;
it should be inserted into the Inogen One G3 as
soon as the dust caps have been removed.
12. Push the column (metal tube) into the device
such that the column (metal tube) is fully seated
into the Inogen One® G3 concentrator. The spring
loaded latch button should fully return to the
closed position.
13. Repeat steps 10-12 to install the other column
(metal tube).
14. Connect the AC power supply cord to the Inogen
One® G3 and plug the power supply AC cord into an
electrical outlet. Do Not Power on the Inogen
One® G3 concentrator.
15. Press and hold the light button for 10 seconds and
screen will display "sieve reset" message. Release
button once message is displayed on screen.
16. Press the bell button once and screen will display "sieve reset successful".
17. Press the power
WARNING Do not use any columns other than those specified in this user manual.
C
!
96-06947-00-01AG3 User Manual Inogen One G3HF-Multilanguage.indb 35
button to turn on the Inogen One® G3, and use normally.
The use of non-specified columns may create a safety hazard and/or
impair equipment performance and will void your warranty.
Closed and locked
Cleaning, Care and Maintenance Chapter 6
35