The specifications and characteristics of your instrument are given in this section. For best
performance, your instrument should be placed where there is ample work space with medium or
subdued illumination and no drafts. The operating conditions (temperature and humidity ranges) are
given in the Operating Conditions section below.

Operating Conditions

ColorFlex EZ can be stored in an area with a temperature range of -20°C to 65°C (-5°F to 150°F) for up
to 3 weeks and can be operated under temperature conditions of 10°C to 40°C (50°F to 104°F). For
specification-level performance, the recommended temperature range is 21-28°C (70-82°F). It may be
operated under relative noncondensing humidity conditions of 10% to 90%. Do not leave ColorFlex EZ
in an area where temperature or humidity extremes are possible.

Physical Characteristics

Weight
Dimensions
Communications Interface
RFI Compliance
Safety Compliance
System Power

Conditions of Illumination and Viewing

Light Source
Source UV content
Lamp Life
Geometry
Detector
Port Diameters/Sample View
Diameters
ColorFlex EZ Supplemental Manual for EasyMatch QC

ColorFlex EZ Specifications

4.5 kg (9.9 lb.)
Height: 16 cm (6.3 in)
Width: 13 cm (5.1 in)
Depth: 36 cm (14.2 in)
3 USB 2.0 ports
FCC Class A (Commercial), IEC, or equivalent
UL, CSA, IEC, or equivalent
100 to 240 VAC, 47 to 63 Hz
Pulsed xenon
Match to D65 with CIE rating of CC or better
>1 million flashes
Directional annular
45° Illumination/0° Viewing
Sealed optics; 256-element scanned array & high-
resolution concave holographic grating
31.8 mm (1.25 in)/25 mm (1.0 in)