Electrical Data; Power And Load Requirements - GE Hart Druck RTX 1000H Series User Manual

Table of Contents

Advertisement

3.9 Electrical Data

Note: If the equipment is certified for use in a hazardous area, refer to
the publication "Calibration data and instructions" for additional
data.
WARNING: Switch off and isolate the power supply before connecting or
disconnecting the transmitter.
CAUTIONS:
1.
The transmitter uses DC power in a 2-wire system to control current
through a resistive load.
2.
Do not apply more than 35 Volts to the loop circuit. The transmitter
may be damaged.
General
The electrical installation must comply with local wiring codes and standards.
To get the full performance from the transmitter, carefully choose the wiring
scheme to be used and take care connecting the transmitter.
Power and maximum load (Figure 3-5)
The total loop resistance must include the connection wire resistance.
Loop DC Power - Volts
Figure 3-5: Power and load requirements
RTX 1000H series user manual
Operating
region
No HART®
Operation
3-7
Note:
For HART® operation,
the minimum loop
resistance is 250 Ω.
K0297 Issue 2